Assessment of the St. Jude Medical Regent prosthetic valve by continuous-wave Doppler and dobutamine stress echocardiography.

Akira Sezai1, Yuji Kasamaki, Keisuke Abe, Mitsumasa Hata, Hisakuni Sekino, Kazuma Shimura, Kazutomo Minami.   

Abstract

BACKGROUND: The St. Jude Medical (SJM) Regent prosthetic valve (St. Paul, MN), a recently developed mechanical valve, is an improvement on the conventional SJM valve, having a wider valve area than the SJM HP valve. We evaluated this mechanical valve by Doppler echocardiography and dobutamine stress echocardiography (DSE).
METHODS: The functions of the SJM Regent valve were evaluated by continuous-wave Doppler echocardiography and DES in 58 cases of aortic valve replacement during a stable postoperative period.
RESULTS: The peak pressure gradient of the replaced valves sized 17, 19, 21, and 23 mm was 27.5 +/- 11.1, 20.0 +/- 9.8, 15.6 +/- 5.7, and 14.3 +/- 9.1, mm Hg respectively, and the effective orifice area index was 0.97 +/- 0.32, 1.01 +/- 0.29, 1.09 +/- 0.30, and 1.41 +/- 0.54 cm(2)/m(2), respectively, with prosthesis-patient mismatch (PPM) found in 1, 3, 2, and 0 cases for the 17-, 19-, 21- and 23-mm valves, respectively, with a total incidence of 10.3%. In 20 cases, the peak pressure gradient and the effective orifice area index were significantly increased during DSE compared with those at rest.
CONCLUSIONS: Although the PPM incidence was 6.6%, it was deemed from the data of DSE and clinical symptoms that there were no clinical issues for such cases of PPM in the early and intermediate phases after operation. Particularly, the effectiveness of the 17- and 19-mm valves in patients with a small aortic annulus was demonstrated, confirming the satisfactory functions of the SJM Regent prosthetic valve. 2010 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20103211     DOI: 10.1016/j.athoracsur.2009.09.013

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  3 in total

Review 1.  Selection of prosthetic valve and evidence--need for the development of Japan's own guidelines.

Authors:  Akira Sezai; Motomi Shiono
Journal:  Ann Thorac Cardiovasc Surg       Date:  2015-06-09       Impact factor: 1.520

2.  Impact of obesity on long-term survival after aortic valve replacement with a small prosthesis.

Authors:  Biao Wang; Hongyang Yang; Tao Wang; Xiquan Zhang; Wenjie Zhu; Guangqing Cao; Shuming Wu
Journal:  Interact Cardiovasc Thorac Surg       Date:  2013-03-25

3.  Application of Regent mechanical valve in patients with small aortic annulus: 3-year follow-up.

Authors:  Dong Zhao; Chunsheng Wang; Tao Hong; Cuizhen Pan; Changfa Guo
Journal:  J Cardiothorac Surg       Date:  2012-09-21       Impact factor: 1.637

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.