Literature DB >> 2010249

ECG diagnostic errors due to improper connection of the right arm and leg cables.

A Guijarro-Morales1, B Gil-Extremera, A Maldonado-Martín.   

Abstract

Two patients were erroneously diagnosed as having acute myocardial infarction or accelerated junctional rhythm due to improper connection of the right arm and right leg cables. In both cases the error led to unsuitable modifications in treatment. A key sign for identifying the technical oversight is minimal voltage in DII in all P, QRS and T waves (dwarf complexes).

Entities:  

Mesh:

Year:  1991        PMID: 2010249     DOI: 10.1016/0167-5273(91)90103-v

Source DB:  PubMed          Journal:  Int J Cardiol        ISSN: 0167-5273            Impact factor:   4.164