| Literature DB >> 2010249 |
A Guijarro-Morales1, B Gil-Extremera, A Maldonado-Martín.
Abstract
Two patients were erroneously diagnosed as having acute myocardial infarction or accelerated junctional rhythm due to improper connection of the right arm and right leg cables. In both cases the error led to unsuitable modifications in treatment. A key sign for identifying the technical oversight is minimal voltage in DII in all P, QRS and T waves (dwarf complexes).Entities:
Mesh:
Year: 1991 PMID: 2010249 DOI: 10.1016/0167-5273(91)90103-v
Source DB: PubMed Journal: Int J Cardiol ISSN: 0167-5273 Impact factor: 4.164