Early Doppler-echocardiography evaluation of 597 prosthetic aortic valves.

Giovanni Minardi1, Paolo G Pino, Carla C Manzara, Giovanni Pulignano, Giuseppe N Viceconte, Giulio G Stefanini, Carlo Gaudio, Francesco Musumeci.   

Abstract

OBJECTIVES: This study was designed to describe the Doppler-echocardiography normal values in the early postoperative phase for Carpentier-Edwards Perimount Aortic Heart Valve bioprosthesis (CEP), St. Jude Medical (SJM) or Medical Regent (SJMR) Mechanical Heart Valves, evaluated by a single experienced echo-laboratory.
METHODS: Five hundred and ninety-seven consecutive patients in our hospital, who had had a CEP, a SJM or SJMR-17 mm implanted due to aortic stenosis, underwent a control Doppler-echocardiography evaluation 4-7 days after surgery. Hemodynamic performance of SJM, SJMR and CEP were accurately described, evaluating flow-dependent (trans-prosthetic velocities and gradients) and flow-independent (effective orifice area, indexed effective orifice area and Doppler velocity index) Doppler-echocardiography parameters.
RESULTS: Of the 597 patients 50.6% were women (n = 302). Mean age was 66.3 +/- 11.7 years. Mean body surface area (BSA) was 1.76 +/- 0.22 m. Mean ejection fraction was 55.3 +/- 10.3%. Two hundred and sixty CEPs and 337 St. Jude mechanical valves (301 SJM and 36 SJMR size-17) were implanted. Comparing size-by-size SJM to CEP, the former were basically less flow obstructive.
CONCLUSION: Our data confirm the wide range of variability, pointing out the need to perform routinely an accurate baseline Doppler-echocardiography evaluation of the hemodynamic profile of prosthetic aortic valves, including flow-dependent and independent parameters, to allow an adequate interpretation at follow-up.

Entities:  

Mesh:

Year:  2010        PMID: 20090552     DOI: 10.2459/JCM.0b013e32832ffe4b

Source DB:  PubMed          Journal:  J Cardiovasc Med (Hagerstown)        ISSN: 1558-2027            Impact factor:   2.160


  1 in total

1.  Early Doppler-echocardiography evaluation of Carpentier-Edwards Standard and Carpentier-Edwards Magna aortic prosthetic valve: comparison of hemodynamic performance.

Authors:  Giovanni Minardi; Giovanni Pulignano; Donatella Del Sindaco; Martina Sordi; Herribert Pavaci; Amedeo Pergolini; Giordano Zampi; Francesca Moschella Orsini; Carlo Gaudio; Francesco Musumeci
Journal:  Cardiovasc Ultrasound       Date:  2011-11-24       Impact factor: 2.062