Radio frequency shielding for a linac-MRI system.

M Lamey1, B Burke, E Blosser, S Rathee, N De Zanche, B G Fallone.   

Abstract

The real-time operation of a linac-MRI system will require proper radio frequency (RF) shielding such that the MRI images can be acquired without extraneous RF noise from the linac. We report on the steps taken to successfully shield the linac from the MRI such that the two devices can operate independently of one another. RF power density levels are reported internally and externally to the RF cage which houses the linac and MRI. The shielding effectiveness of the RF cage has been measured in the frequency range 1-50 MHz and is presented. Lastly MRI images of two phantoms are presented during linac operation. This work illustrates that the accelerating structure of a linac and an MRI can be housed within the same RF cage. The 6 MV linac can be operated to produce radiation with no measurable degradation in image quality due to RF effects.

Mesh:

Year:  2010        PMID: 20090181     DOI: 10.1088/0031-9155/55/4/006

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

Review 1.  MR-guided radiation therapy: transformative technology and its role in the central nervous system.

Authors:  Yue Cao; Chia-Lin Tseng; James M Balter; Feifei Teng; Hemant A Parmar; Arjun Sahgal
Journal:  Neuro Oncol       Date:  2017-04-01       Impact factor: 12.300

2.  Performance investigation of LabPET II detector technology in an MRI-like environment.

Authors:  Narjes Moghadam; Louis Arpin; Romain Espagnet; Jonathan Bouchard; Nicolas Viscogliosi; Roger Lecomte; Réjean Fontaine
Journal:  Phys Med Biol       Date:  2020-01-27       Impact factor: 3.609

3.  Compact bunker shielding assessment for 1.5 T MR-Linac.

Authors:  Jiwon Sung; Yeonho Choi; Jun Won Kim; Ik Jae Lee; Ho Lee
Journal:  Sci Rep       Date:  2022-04-25       Impact factor: 4.996

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.