Simultaneous intraoperative measurement of cardiac output by thermodilution and transtracheal Doppler.

L C Siegel1, D C Fitzgerald, R H Engstrom.   

Abstract

Intraoperative measurement of cardiac output with transtracheal Doppler (DOP) was compared with that measured by thermodilution (TD). Cardiac output was measured simultaneously with both methods in 17 adult patients. For 86 pairs of measurements, the average difference between the two techniques was -0.21.min-1. This bias had a standard deviation of 1.71.min-1. The average of the absolute value of the difference between measurements made with the two techniques was 1.31.min-1, with a standard deviation of 1.11.min-1. The limits of agreement were -3.6 to 3.11.min-1. Linear regression yielded the following equation: DOP = 0.62 TD + 1.54 l.min-1 (r = 0.63). To evaluate the ability of transtracheal DOP to trend changes in cardiac output, the changes in cardiac output at sequential time points were compared for the two techniques. The average difference in the changes in cardiac output measured by the two techniques was 0.01.min-1. This bias had a standard deviation of 1.71.min-1. In conclusion, the transtracheal DOP technique did not reproduce the measurement of cardiac output by TD intraoperatively. Transtracheal DOP did not accurately trend changes in the TD measurement. These findings were obtained from patients with cardiovascular disease, and the conclusions may depend in part on the patient population and the investigators' experience with the transtracheal DOP technique.

Entities:  

Mesh:

Year:  1991        PMID: 2008947     DOI: 10.1097/00000542-199104000-00008

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  3 in total

1.  A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques.

Authors:  L A Critchley; J A Critchley
Journal:  J Clin Monit Comput       Date:  1999-02       Impact factor: 2.502

Review 2.  Monitoring during paediatric cardiac anaesthesia.

Authors:  J P Purday
Journal:  Can J Anaesth       Date:  1994-09       Impact factor: 5.063

Review 3.  Cardiac output monitoring by echocardiography: should we pass on Swan-Ganz catheters?

Authors:  A C Perrino
Journal:  Yale J Biol Med       Date:  1993 Sep-Oct
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.