Quantitative three-dimensional power Doppler angiography: a flow-free phantom experiment to evaluate the relationship between color gain, depth and signal artifact.

W P Martins1, N J Raine-Fenning, R A Ferriani, C O Nastri.   

Abstract

OBJECTIVES: To evaluate the presence of false flow three-dimensional (3D) power Doppler signals in 'flow-free' models.
METHODS: 3D power Doppler datasets were acquired from three different flow-free phantoms (muscle, air and water) with two different transducers and Virtual Organ Computer-aided AnaLysis was used to generate a sphere that was serially applied through the 3D dataset. The vascularization flow index was used to compare artifactual signals at different depths (from 0 to 6 cm) within the different phantoms and at different gain and pulse repetition frequency (PRF) settings.
RESULTS: Artifactual Doppler signals were seen in all phantoms despite these being flow-free. The pattern was very similar and the degree of artifact appeared to be dependent on the gain and distance from the transducer. False signals were more evident in the far field and increased as the gain was increased, with false signals first appearing with a gain of 1 dB in the air and muscle phantoms. False signals were seen at a lower gain with the water phantom (-15 dB) and these were associated with vertical lines of Doppler artifact that were related to PRF, and disappeared when reflections were attenuated.
CONCLUSIONS: Artifactual Doppler signals are seen in flow-free phantoms and are related to the gain settings and the distance from the transducer. In the in-vivo situation, the lowest gain settings that allow the detection of blood flow and adequate definition of vessel architecture should be used, which invariably means using a setting near or below the middle of the range available. Additionally, observers should be aware of vertical lines when evaluating cystic or liquid-containing structures. (c) 2010 ISUOG. Published by John Wiley & Sons, Ltd.

Entities:  

Mesh:

Year:  2010        PMID: 20084642     DOI: 10.1002/uog.7562

Source DB:  PubMed          Journal:  Ultrasound Obstet Gynecol        ISSN: 0960-7692            Impact factor:   7.299


  10 in total

Review 1.  Three-dimensional ultrasound as a predictor of pregnancy in patients undergoing ART.

Authors:  Cemil Yaman; Richard Mayer
Journal:  J Turk Ger Gynecol Assoc       Date:  2012-06-01

2.  Smart Blood Vessel Detection System for Laparoscopic Surgery.

Authors:  Ching-Chia Li; Bor-Shing Lin; Sheng-Chen Wen; Yuan-Teng Liang; Hung-Yu Sung; Jhen-Hao Jhan; Bor-Shyh Lin
Journal:  IEEE J Transl Eng Health Med       Date:  2022-03-11       Impact factor: 3.316

3.  Applying spatial-temporal image correlation to the fetal kidney: Repeatability of 3D segmentation and volumetric impedance indices.

Authors:  Bonita Gu; Gordon N Stevenson; Ana Ferreira; Sudeshni Pathirana; Jennifer Sanderson; Amanda Henry; Jennifer Alphonse; Alec W Welsh
Journal:  Australas J Ultrasound Med       Date:  2018-05-11

4.  Comparative analysis of placental vasculature and placental volume in normal and IUGR pregnancies with the use of three-dimensional Power Doppler.

Authors:  Michal Pomorski; Mariusz Zimmer; Jerzy Florjanski; Joanna Michniewicz; Artur Wiatrowski; Tomasz Fuchs; Ewa Milnerowicz-Nabzdyk
Journal:  Arch Gynecol Obstet       Date:  2011-07-09       Impact factor: 2.344

5.  Placental vascularization alterations in hypertensive disorders complicating pregnancy (HDCP) and small for gestational age with HDCP using three-dimensional power doppler in a prospective case control study.

Authors:  Ting Yuan; Ting Zhang; Zhen Han
Journal:  BMC Pregnancy Childbirth       Date:  2015-10-05       Impact factor: 3.007

6.  Reduced placental volume and flow in severe growth restricted fetuses.

Authors:  Renata Montes Dourado Abulé; Lisandra Stein Bernardes; Giovana Farina Doro; Seizo Miyadahira; Rossana Pulcinelli Vieira Francisco
Journal:  Clinics (Sao Paulo)       Date:  2016-07       Impact factor: 2.365

7.  Short-Term Exposure to Urban Air Pollution and Influences on Placental Vascularization Indexes.

Authors:  Karen Hettfleisch; Lisandra Stein Bernardes; Mariana Azevedo Carvalho; Luciana Duzolina Manfré Pastro; Sandra Elisabete Vieira; Silvia R D M Saldiva; Paulo Saldiva; Rossana Pulcineli Vieira Francisco
Journal:  Environ Health Perspect       Date:  2016-07-06       Impact factor: 9.031

8.  Three-dimensional power Doppler ultrasound evaluation of placental blood flow in normal monochorionic diamniotic twin pregnancies.

Authors:  Wei Sun; Shaowei Yin; Qiuju Wei; Ying Zhang; Zeyu Yang; Ailu Cai; Yu Wang; Wenjia Lei
Journal:  BMC Pregnancy Childbirth       Date:  2018-11-14       Impact factor: 3.007

Review 9.  How to use power Doppler ultrasound in transvaginal assessment of uterine fibroids.

Authors:  M Frijlingh; L Juffermans; R de Leeuw; C de Bruyn; D Timmerman; T van den Bosch; J A F Huirne
Journal:  Ultrasound Obstet Gynecol       Date:  2022-08       Impact factor: 8.678

10.  4D Doppler Ultrasound in High Grade Serous Ovarian Cancer Vascularity Evaluation-Preliminary Study.

Authors:  Marek Jerzy Kudla; Michal Zikan; Daniela Fischerova; Mateusz Stolecki; Juan Luis Alcazar
Journal:  Diagnostics (Basel)       Date:  2021-03-24
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.