A rabbit model for efficacy evaluation of endovascular coil materials.

Masaki Ebara1, Ichiro Yuki, Yuichi Murayama, Takayuki Saguchi, Yih-Lin Nien, Harry V Vinters, Fernando Vinuela, Toshiaki Abe.   

Abstract

BACKGROUND: To investigate biomaterials seeking for their possible use for aneurysm treatment, in vivo screening tests using a number of potential materials are required. However, there is no established animal model that is suitable for such purpose. Some models require special preparation of tested materials for transcatheter delivery and others are inappropriate in view of their cost-effectiveness. The purpose of this study is to establish an animal model that overcomes these limitations and help us select potential materials before the preclinical evaluation.
METHODS: Bilateral CCAs in a rabbit were surgically ligated, and a 2-cm segment of either a bare platinum coil or a polymeric coil (a platinum coil coated with PLGA 10/90) was implanted into each blind-ended arterial segment (n = 26). They were harvested at day 1, 7, 10, 14, or 30, respectively. Angiographic and histologic evaluations as well as quantitative analysis on the development of the organized thrombus were performed.
RESULTS: One day after the implantation, both platinum and PLGA coils were surrounded by immature thrombus that was induced by blood flow stagnation in the arterial segment. At day 7, minimal thrombus organization was observed around both types of materials. At postimplantation days 10 and 14, fibrocellular responses, the early findings of the thrombus organization process, were observed in both material groups. Such histologic findings were more prominent in the PLGA coil group as compared to the platinum coil group (day 10, P = .051; day 14, P = .011). Well-organized thrombus was observed in both material groups at day 30 without showing statistical difference (P = .12).
CONCLUSION: Given the cost-effectiveness, the simple material preparation process, and its feasible histologic evaluation methods, this new animal model can be useful in screening other potential biomaterials for the development of new coil devices. Copyright 2009 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 20082832     DOI: 10.1016/j.surneu.2009.04.017

Source DB:  PubMed          Journal:  Surg Neurol        ISSN: 0090-3019


  2 in total

1.  Intravascular tissue reactions induced by various types of bioabsorbable polymeric materials: correlation between the degradation profiles and corresponding tissue reactions.

Authors:  Ichiro Yuki; Naoyuki Uchiyama; Yuichi Murayama; Yih-Lin Nien; Daniel Lee; Masaki Ebara; Akira Ishii; Alexander Chiang; Harry V Vinters; Ichiro Nishimura; Benjamin M Wu; Fernando Vinuela
Journal:  Neuroradiology       Date:  2010-02-10       Impact factor: 2.804

2.  Preclinical extracranial aneurysm models for the study and treatment of brain aneurysms: A systematic review.

Authors:  Serge Marbacher; Fabio Strange; Juhana Frösén; Javier Fandino
Journal:  J Cereb Blood Flow Metab       Date:  2020-03-03       Impact factor: 6.200

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.