Minimizing ferromagnetic artefact with metallic lumbar total disc arthroplasty devices at adjacent segments: technical note.

Laurence A G Marshman1, Graeme Strong, Matthew Trewhella, Ata Kasis, Tai Friesem.   

Abstract

STUDY
DESIGN: Technical report.
OBJECTIVE: To minimize the gross artefact associated with a conventional metallic lumbar total disc arthroplasty (TDA) device on a conventional high field-strength (1.5 Tesla [T]) magnetic resonance imaging (MRI) scanner. SUMMARY OF BACKGROUND DATA: Gross artefact is often apparent with ferromagnetic metallic TDA devices on conventional high field-strength MRI scanners. Such artefact completely obliterates MRI assessment at the operated level, and usually obscures visualization of adjacent segments. Because adjacent segment preservation is a raison d'être of spinal TDA, clarity of imaging at this latter level is imperative. A failure to image adjacent segments may presage investigations, which are either invasive (e.g., myelography) or associated with significant radiation hazard (e.g., computed tomography), both with significantly less diagnostic sensitivity. This could negatively direct TDA choice with certain TDAs.
METHODS: Progressive modifications to specific imaging parameter settings were sought on a conventional high field-strength (1.5T) closed-bore scanner to match the minimal artefact previously observed on a lower field-strength (0.3T) open scanner. Direct comparisons were made between each modified protocol image obtained initially using a phantom; however, routine postoperative MRIs were subsequently obtained in n = 40 patients following lumbar TDA insertion.
RESULTS: Key parameter modifications were required in the receive bandwidth, the strength of the frequency encoding gradient, as well as in the echo train length. The use of higher specification "focused gradients" was also avoided. The overall effect was to reduce the slew rate of the gradients, which limited artefact due to a decrease in phase dispersion. Such appearances effectively matched with those previously obtained on the low field-strength (0.3T) open scanner in n = 40 patients.
CONCLUSION: Relatively simple modifications to MRI parameter settings can be made on conventional high field-strength (1.5T) closed-bore scanners, which minimize metal artefact and enhance imaging of adjacent segments with ferromagnetic TDA devices. Such modifications effectively match appearances to those obtained with outmoded low field-strength (0.3T) open-bore scanners.

Entities:  

Mesh:

Year:  2010        PMID: 20081522     DOI: 10.1097/BRS.0b013e3181c838cc

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  3 in total

1.  Facet arthropathy evaluation: CT or MRI?

Authors:  Linda Berg; Hanne Thoresen; Gesche Neckelmann; Håvard Furunes; Christian Hellum; Ansgar Espeland
Journal:  Eur Radiol       Date:  2019-02-22       Impact factor: 5.315

2.  We Need to Talk about Lumbar Total Disc Replacement.

Authors:  Stephen Beatty
Journal:  Int J Spine Surg       Date:  2018-08-03

3.  Can we measure the spiral and uterine artery blood flow by real-time sonography and Doppler indices to predict spontaneous miscarriage in a normal-risk population?

Authors:  Mehmet Burak Özkan; Elif Ozyazici; Baris Emiroglu; Enis Özkara
Journal:  Australas J Ultrasound Med       Date:  2015-12-31
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.