Clinical evaluation of a new chest tube used in neonates.

A L Jung1, J Nelson, M B Jenkins, W A Hodson.   

Abstract

We report preliminary experience with a newly designed chest tube (JCT), for evacuation of neonatal pneumothorax. The catheter has a unique pigtail confirguration at the distal end, intended to simplify placement and minimize chest wall and lung trauma by reduced tube size and depth and insertion. Thirty-eight JCTs were placed in neonates with pneumothoraces. Neonates' birth weights ranged from 400 to 3,595 grams. All 38 tubes immediately relieved clinical signs of pneumothoraces. Thirty-five (92%) tubes immediately fully evacuated the pneumothoraces as evidences on chest radiograph. Twelve pneumothoraces partially reoccurred at a mean of 24 hours following JCT placement. These tubes were either irrigated or replaced. This newly configured chest tube functions effectively in the treatment of neonatal pneumothorax.

Entities:  

Mesh:

Year:  1991        PMID: 2007311     DOI: 10.1177/000992289103000204

Source DB:  PubMed          Journal:  Clin Pediatr (Phila)        ISSN: 0009-9228            Impact factor:   1.168


  1 in total

1.  Neonatal chest drain insertion--an animal model.

Authors:  J O Hourihane; P A Crawshaw; M A Hall
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  1995-03       Impact factor: 5.747