| Literature DB >> 2007311 |
A L Jung1, J Nelson, M B Jenkins, W A Hodson.
Abstract
We report preliminary experience with a newly designed chest tube (JCT), for evacuation of neonatal pneumothorax. The catheter has a unique pigtail confirguration at the distal end, intended to simplify placement and minimize chest wall and lung trauma by reduced tube size and depth and insertion. Thirty-eight JCTs were placed in neonates with pneumothoraces. Neonates' birth weights ranged from 400 to 3,595 grams. All 38 tubes immediately relieved clinical signs of pneumothoraces. Thirty-five (92%) tubes immediately fully evacuated the pneumothoraces as evidences on chest radiograph. Twelve pneumothoraces partially reoccurred at a mean of 24 hours following JCT placement. These tubes were either irrigated or replaced. This newly configured chest tube functions effectively in the treatment of neonatal pneumothorax.Entities:
Mesh:
Year: 1991 PMID: 2007311 DOI: 10.1177/000992289103000204
Source DB: PubMed Journal: Clin Pediatr (Phila) ISSN: 0009-9228 Impact factor: 1.168