Literature DB >> 20068369

Fluid temperature at the corneal endothelium during phacoemulsification: comparison of an ophthalmic viscosurgical device and balanced salt solution using the finite element method.

Somporn Reepolmaha1, Wiroj Limtrakarn, Wanlaya Uthaisang-Tanechpongtamb, Pramote Dechaumphai.   

Abstract

PURPOSE: The purpose of this study was to estimate and compare the temperatures of two different anterior chamber solutions at the corneal endothelial level during phacoemulsification. An ophthalmic viscosurgical device (OVD) and balanced salt solution (BSS) were compared using the finite element method (FEM).
MATERIALS AND METHODS: The thermal properties of an OVD (IAL-F) and BSS were studied in an experimental setting. A computer-aided design model of ocular anatomy was created in two dimensions. The phaco needle was considered to be the only source of heat generation. Then, the FEM was used to demonstrate the transient temperature distribution in the two ocular models at 10, 20, 30, 40, 50 and 60 s. In these models, the anterior chamber was filled with IAL-F (IAL-F model) or BSS (BSS model).
RESULTS: The heat generation rate of the phaco needle was 0.0004 cal/s/mm(2). The maximum corneal endothelial temperatures for the two models at 60 s were 52.67 and 41.57 degrees C, respectively. The experimental IAL-F model showed fewer changes in temperature for any given time and location. At larger distances from the heat source, less temperature variation was detected.
CONCLUSION: Phacoemulsification is a potential heat-generating procedure performed between the delicate anterior chamber structures. During this procedure, IAL-F protects the endothelium against heat better than BSS. Copyright 2009 S. Karger AG, Basel.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 20068369     DOI: 10.1159/000272020

Source DB:  PubMed          Journal:  Ophthalmic Res        ISSN: 0030-3747            Impact factor:   2.892


  4 in total

1.  Comparison of surgical parameters using different lens fragmentation patterns in eyes undergoing laser-assisted cataract surgery.

Authors:  Harvey S Uy; Pik Sha Chan; Raquel Gil-Cazorla; Sunil Shah
Journal:  Int Ophthalmol       Date:  2019-03-04       Impact factor: 2.031

2.  Effects of phacoemulsification with versus without viscoelastic devices on surgical outcomes.

Authors:  İbrahim Taşkın; Lokman Aslan
Journal:  Int Ophthalmol       Date:  2017-05-09       Impact factor: 2.031

3.  Active-fluidics versus gravity-fluidics system in phacoemulsification for age-related cataract (AGSPC): study protocol for a prospective, randomised, double-blind, controlled clinical trial.

Authors:  Yu Luo; Hongyu Li; Wenqian Chen; Yi Gao; Tianju Ma; Zi Ye; Zhaohui Li
Journal:  BMJ Open       Date:  2022-01-20       Impact factor: 2.692

4.  Comparison of cumulative dispersed energy between conventional phacoemulsification and femtosecond laser-assisted cataract surgery with two different lens fragmentation patterns.

Authors:  Hung-Yuan Lin; Shu-Ting Kao; Ya-Jung Chuang; Shuan Chen; Pi-Jung Lin
Journal:  Lasers Med Sci       Date:  2021-04-19       Impact factor: 3.161

  4 in total