Literature DB >> 20055343

Incomplete seating of a metal-backed alumina liner in ceramic-on-ceramic total hip arthroplasty.

Jaime A Carvajal Alba1, Eric D Schiffman, Sean P Scully, Hari K Parvataneni.   

Abstract

Metal-backed modular ceramic bearing systems using a recessed alumina liner in a titanium sleeve were developed to decrease ceramic chipping or fracture due to femoral neck impingement after total hip arthroplasty (THA). However, malseating of the metal-backed ceramic liner has recently been described. The goal of this study was to assess the prevalence, etiology, and clinical relevance of this event. Between 2005 and 2008, 51 consecutive patients (61 hips) underwent THA with a metal-backed alumina liner housed in a titanium shell. The metal-backed ceramic liner was aligned, seated, and impacted into the shell, and satisfaction in terms of liner stability and seating was confirmed intraoperatively. Postoperative assessment of seating was assessed with standard radiographs. Liner seating was classified as well seated, suspicious, or malseated. Seven liners (11.5%) were found to be malseated and 4 (6.5%) were considered suspicious. Radiographically, there was a gap between the liner and the shell located inferomedially in 4 patients and superolaterally in 3 patients. Two liners subsequently seated at 1 and 3 months postoperatively, respectively. No dislodgement, failures, or adverse events were identified. There were no revision surgeries. The significant percentage of malseated liners were potentially attributed to poor exposure, bony/soft tissue interposition, and surgeon learning curve. Copyright 2010, SLACK Incorporated.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20055343     DOI: 10.3928/01477447-20091124-11

Source DB:  PubMed          Journal:  Orthopedics        ISSN: 0147-7447            Impact factor:   1.390


  4 in total

1.  Preventing ceramic liner fracture after Delta ceramic-on-ceramic total hip arthroplasty.

Authors:  Young-Kyun Lee; Jae-Young Lim; Yong-Chan Ha; Tae-Young Kim; Woon-Hwa Jung; Kyung-Hoi Koo
Journal:  Arch Orthop Trauma Surg       Date:  2020-06-12       Impact factor: 3.067

2.  Canted seating of the Stryker Modular Dual Mobility liner within a Trident hemispherical acetabular shell.

Authors:  Scott M Eskildsen; Erik C Olsson; Daniel J Del Gaizo
Journal:  Arthroplast Today       Date:  2016-01-06

3.  Use of a modular hip dual-mobility articulation in patients with high risk of dislocation: a relatively small-sized acetabulum in Asian patients may limit its use.

Authors:  Ping Keung Chan; Sum Lik Cheung; Kar Hei Lam; Wing Chiu Fung; Vincent Wai Kwan Chan; Amy Cheung; Man Hong Cheung; Henry Fu; Chun Hoi Yan; Kwong Yuen Chiu
Journal:  Arthroplasty       Date:  2021-05-03

4.  Spontaneous resolution of asymptomatic alumina matrix composite ceramic liner dissociation: a case report.

Authors:  He Xiao; Jian Wang; Nian-Ye Zheng; Zhan-Jun Shi
Journal:  BMC Musculoskelet Disord       Date:  2022-08-20       Impact factor: 2.562

  4 in total