| Literature DB >> 20052210 |
G Klatt1, R Gebs, C Janke, T Dekorsy, A Bartels.
Abstract
We report a terahertz time-domain spectrometer with more than 6 THz spectral coverage and 1 GHz resolution based on high-speed asynchronous optical sampling. It operates at 2 kHz scan rate without mechanical delay stage. The frequency error of the system at 60 s acquisition time is determined by comparing a measured water vapor absorption spectrum to data reported in the HITRAN database. The mean error of 87 evaluated absorption lines is 142 MHz.Entities:
Mesh:
Year: 2009 PMID: 20052210 DOI: 10.1364/OE.17.022847
Source DB: PubMed Journal: Opt Express ISSN: 1094-4087 Impact factor: 3.894