Literature DB >> 20049575

Accidental fracture of the tip of the Coopdech bronchial blocker during insertion for one lung ventilation.

Arun Venkataraju1, Christopher Rozario, Palanikumar Saravanan.   

Abstract

PURPOSE: The distal tip of a Coopdech bronchial blocker has a preformed angulation to aid placement in the desired bronchus. We report two cases wherein this design may have resulted in distal tip fracture due to entanglement at the level of the Murphy's eye of the endotracheal tube or at the carina. CLINICAL FEATURES: A 49-yr-old female had a Coopdech bronchial blocker inserted into her right main bronchus for video-assisted thoracoscopic (VAT) lung biopsy. Resistance was encountered on its insertion, followed by confirmation of its position by fibreoptic bronchoscopy. As lung isolation was inadequate, bronchoscopy was repeated during surgery. This showed fracture of the blocker tip that required patient repositioning and insertion of another blocker. In a second incident, a bronchial blocker was inserted into the right main bronchus of a 19-yr-old male for VAT bullectomy. This procedure was performed under continuous fibreoptic guidance. Nevertheless, it was difficult to pass the blocker tip beyond the Murphy's eye of the endotracheal tube, as repeated attempts resulted in its entanglement and fracture. Another blocker was inserted by maneuvering the tip beyond the Murphy's eye.
CONCLUSION: The preformed tip of the Coopdech bronchial blocker can be damaged at the Murphy's eye of the endotracheal tube or at the carina. This can result in tip fracture, especially during insertion into the right main bronchus. Maneuvering the tip away from the Murphy's eye can circumvent this problem. Continuous bronchoscopic guidance should be used as recommended by the manufacturer.

Entities:  

Mesh:

Year:  2010        PMID: 20049575     DOI: 10.1007/s12630-009-9261-0

Source DB:  PubMed          Journal:  Can J Anaesth        ISSN: 0832-610X            Impact factor:   5.063


  5 in total

1.  Rigid Bronchoscopic Placement of Fogarty Catheter as a Bronchial Blocker for One Lung Isolation and Ventilation in Infants and Children Undergoing Thoracic Surgery: A Single Institution Experience of 27 Cases.

Authors:  Sunil Kant Kamra; Ashwin Ashok Jaiswal; Amrish Kumar Garg; Manoj Kumar Mohanty
Journal:  Indian J Otolaryngol Head Neck Surg       Date:  2016-10-15

2.  Bronchial blocker for one-lung ventilation: An unanticipated complication.

Authors:  Namrata Niwal; Priya Ranganathan; Jigeeshu Divatia
Journal:  Indian J Anaesth       Date:  2011-11

3.  The tip fracture of the Coopdech bronchial blocker during insertion in the patient with lung surgery.

Authors:  Deokkyu Kim; Ji-Seon Son; Hyungsun Lim; Jun Ho Lee; Eun Joo Jang
Journal:  Korean J Anesthesiol       Date:  2014-12

4.  Ultrasound-Guided Selective Bronchial Intubation: A Feasibility Study in Pediatric Animal Model.

Authors:  Sara Hora Gomes; Alice Miranda; José Miguel Pêgo; Patrício S Costa; Jorge Correia-Pinto
Journal:  Front Med (Lausanne)       Date:  2022-06-15

5.  A randomised comparison of the efficacy of a Coopdech bronchial blocker and a double-lumen endotracheal tube for minimally invasive esophagectomy.

Authors:  Tian-Hua Zhang; Xiao-Qing Liu; Long-Hui Cao; Jian-Hua Fu; Wen-Qian Lin
Journal:  Transl Cancer Res       Date:  2020-08       Impact factor: 1.241

  5 in total