Literature DB >> 20046478

Monomer Release from Resin Based Dental Materials Cured With LED and Halogen Lights.

Asli Topaloglu Ak1, A Riza Alpoz, Oguz Bayraktar, Fahinur Ertugrul.   

Abstract

OBJECTIVES: To measure the release of TEGDMA and BisGMA from two commercially available composite resins; Filtek Z 250 (3M ESPE, Germany), Leaddent (Leaddent, Germany) and two fissure sealants; Helioseal F (3M ESPE, Germany) Enamel Loc (Premiere Rev, USA) over 1, 3 and 7 days after polymerization with standard quartz-tungsten halogen Coltolux II (QHL) (Coltene Switzerland) and a standard blue light emitting diode Elipar Freelight 2 (3M ESPE, Germany).
METHODS: 9 samples of each material were placed in disc shaped specimens in 1 mm of thickness and 10 mm in diameter (n=36). Each material was polymerized using LED for 20 s (n=12), 40 s (n=12) and halogen for 40 s (n=12), respectively. High Performance Liquid Chromatography (HPLC) was used to measure the amount of monomers released over 1, 3 and 7 days. Data was analyzed using one way ANOVA and Bonferroni test for multiple comparisons with a significance level of .05.
RESULTS: LED 20 sec group showed the highest release of monomers at 1, 3 and 7 days in sealant groups. Halogen 40 sec group resulted highest release of monomers for Leaddent at all time intervals (P<.05)
CONCLUSIONS: Efficiency of the curing unit and applying the recommended curing time of the light activated resin based dental materials is very important to protect the patient from potential hazards of residual monomers.

Entities:  

Keywords:  BisGMA; Composite resin; Fissure sealant; Residual monomer; TEGDMA

Year:  2010        PMID: 20046478      PMCID: PMC2798788     

Source DB:  PubMed          Journal:  Eur J Dent


INTRODUCTION

Light activated resin based dental restorative materials are most widely accepted among clinicians for restorations and fissure sealants.1,2 Among their advantages, control of contour during restoration, improved color stability, increased polymerization and rapid setting compared to chemical activated materials come primarily.3 However, despite these advantages; both chemical and light activated dental materials are under scope for their biological effects and biocompatibility.4–7 One of the most common drawbacks of a composite resin and a resin based fissure sealant is the inadequate polymerization which results with high residual monomers. Consequently, this leads to inferior physical properties.8,9 Previous studies indicated that, oxygen prevents polymerization of monomer by the formation of an inhibition zone on the surface of a resin in contact with air.10,11 Hence, sealant materials have even more potential for leaving uncured resin components since these materials are cured without an occlusal matrix.12 These unreacted components may either attach to polymer chain or integrate into host systemically by saliva, mucosal tissue membranes and pulp tissues. This may result with allergic or adverse side effects in both patient and dental staff.13,14 The significance of removing the leaching monomers from the oxygen inhibited layer of resin based materials have been documented in the literature.12,15 Factors affecting the polymerization of the resins include those, directly related to restorative material and light curing units. In order to achieve the adequate polymerization to overcome the residual monomers, curing units and techniques have been studied by many researches.16,17 Halogen lights have been most commonly used devices for the polymerization of resin based dental materials. This low cost technology device have drawbacks such as decline of irradiance over time due to bulb and filter ageing which could lead to inadequate polymerization.18–20 Different technologies for light curing resin based materials have been developed to overcome these disadvantages. One such alternative is light emitting diode (LED) which is a highly efficient light source that produces light within a narrow spectral range. LED is very popular among paediatric dentists particularly, since less chair time and an adequate polymerization is the main goal and there are few studies on the curing performance of different curing units using HPLC (High Performance Liquid Chromatography) in time related elution of term.21–23 The objective of this study was to determine the time related release residual monomers from different brands of commercially available resin based fissure sealants and composite resins polymerized using different curing units with different exposure time.

MATERIALS AND METHODS

Sample fabrication

Two composite resins, Filtek Z 250 (3M ESPE, Seefeld, Germany) and Lead Dent (Leaddent, Germany) and two resin based fissure sealants, Helioseal F (Ivoclar Vivadent AG, Liechtenstein, Germany) and Enamel Loc (Premier Dental Products Company, REV, USA) were used in this study. The chemical compositions of these materials are listed in Table 1.
Table 1

Chemical composition of resin based dental materials used.

Resin Based Dental MaterialsChemical CompositionManufacturer
Helioseal FBis-GMA, TEGDMA,UDMA UDMA, TEGDMA, Methacrylated phosphoric acid esters,Ivoclar Vivadent AG, Liechtenstein, GermanyPremier Dental
Enamel Loc4-methacryloxy ethyltrimellitic acidProducts Company, REV, USA
Filtek Z 250Bis-GMA, UDMA, Bis-EMA3M, ESPE, Seefeld, Germany
LeaddentBis-GMA, TEGDMA, UDMALeadent, Germany
For each resin based dental materials, 9 discshaped specimens (10 mm diameter and 1 mm thickness) were prepared using silicone rubber molds (n=36). The molds were placed on flat glass plates on top of acetate strips and then filled with resin based material. The material was covered with an acetate strip and gently pressed with another glass plate against the mold to extrude excess material.

Curing method and exposure time

Halogen light Coltolux II (QHL) (Coltene, Switzerland) and LED Elipar Freelight 2 (3M ESPE, Germany) were used for the polymerization of the samples. The technical details of halogen light and LED are presented in Table 2. Nine samples of each material were (n=36), polymerized using LED for 20 s (n=12), 40 s (n=12) and halogen light for 40 s (n=12), respectively. The distance between the light source and sample was standardized by using a 1 cm glass plate. The light tip was in close contact with the restoration surface during polymerization. All specimens were prepared in a temperature controlled room at 23±1°C. Immediately after light-curing, the cover glasses were removed from the mold.
Table 2

Technical details of the halogen and LED light-curing units.

Light Curing UnitsWavelengthIntensityManufacturer
Coltolux II400–515 nm400 mW/cm2(Coltene, Switzerland)
LED Elipar Freelight 2430–480 nm1200 mW/cm2(3M ESPE, Germany)

Release studies and HPLC analysis

After curing, samples were immediately submerged in 5 ml of ethanol at 37°C for 7 days and were analyzed at 1, 3 and 7 days, respectively. BisGMA and TEGDMA released from samples were analyzed by reversed phase HPLC high performance liquid chromatography (HPLC) equipped with a diode array detector. The detection was performed at wavelength of 254 nm. The stationary phase was a C18 LiChrospher 100 analytical column (250×4 mm i.d.) with a particle size of 5 mm thermostated at 35°C. The mobile phase was an isocratic solution of 65% acetonitrile, at a flow rate of 1.5 mL min−1. The injection volume was 20 μl and the measurements were performed after 1, 3 and 7 days and repeated 3 times for all release media. BisGMA and TEGDMA monomers were identified by comparison of their retention times with the corresponding standards. Chromatogram of standard monomer sample is given in Figure 1.
Figure 1

Chromatogram of the standard monomer sample.

One-way analysis of variance was used to analyze data on the quantity of monomer released by monomer type, time intervals and different light curing methods. Multiple comparisons were analyzed using Bonferroni test. Statistical analysis was conducted at a significance level of P<.05.

RESULTS

Tables 3–6 show the mean amounts of BisGMA and TEGDMA released from Helioseal F, Enamel Loc, Filtek Z 250 and Leaddent samples respectively after 1, 3 and 7 days following immersion in ethanol. In Helioseal F group, the release of BisGMA was statistically higher in LED 20 s group compared to LED 40 s group for all time intervals (P<.05) whereas this difference was significant for TEGDMA only after 3 days, (P<.05). In Enamel Loc group, LED 20 s released more TEGDMA in comparison to LED 40 s after 1 and 3 days (P<.05). For BisGMA, LED 20 s group released more monomer than LED 40 s and QTH 40 s at all time intervals (P<.05).
Table 3

Mean TEGDMA and BisGMA release (concg/L) from Helioseal F fissure sealant samples in to ethanol after 1,3 and 7 days. a,b,c,d P<.05.

Immersion time (days)TEGDMA (concg/l)BisGMA (concg/l)
LED 20 mean±sd (n=3)LED 40 mean±sd (n=3)QTH 40 mean±sd (n=3)LED 20 mean±sd (n=3)LED 40 mean±sd (n=3)QTH 40 mean±sd (n=3)
17.55±0.326.54±1.06.74±0.660.41a±0.070.24a±0.040.3±0.05
39.61b±0.537.54b±0.848.24±0.70.45c±0.040.31c±0.030.36±0.04
711.76±1.069.98±1.039.62±0.580.49d±0.030.35d±0.050.38±0.05
Table 6

Mean TEGDMA and BisGMA release (concg/L) from Leaddent composite resin samples in to ethanol after 1,3 and 7 days. a,b,c,d,e,f,g,h P<.05.

Immersion time (days)TEGDMA (concg/l)BisGMA (concg/l)
LED 20 mean±sd (n=3)LED 40 mean±sd (n=3)QTH 40 mean±sd (n=3)LED 20 mean±sd (n=3)LED 40 mean±sd (n=3)QTH 40 mean±sd (n=3)
10.15±0.050.10a±0.010.28a±0.080.12±0.050.07b±0.010.23b±0.07
30.21±0.010.11c±0.010.31c±0.090.17±0.010.08d±0.010.27d±0.07
70.23e±0.010.12f±0.010.43e,f±0.090.20g±0.010.09h±0.000.37g,h±0.09
There was no significant difference noted for release of TEGDMA monomer among curing methods at all time intervals in Filtek Z 250 composite resin samples (P>.05). However, there was a significant difference in the release of BisGMA after 3 and 7 days when halogen 40s was compared to LED 40 s group. LED 40 s group released less monomer (P<.05). In Leaddent composite resin sample group, the release of both TEGDMA and BisGMA were significantly less for LED 40 s group compared to halogen 40 s for all time intervals (P<.05). This difference was significant compared to LED 20 s only after 7 days.

DISCUSSION

It has been shown that various components may be released from resin based dental restorative materials into the oral environment.9 These components namely, degradation products, oligomers and residual monomers can affect the biocompatibility of the materials. The amount of residual, unreacted monomer in a resin based dental material has been implicated in causing cytotoxicity and leading to a mechanical degradation in the material itself. Recently it has been also reported that released monomers from dental materials show synthetic eustrogenic activity.10,24 Hence researches have focused on the contributing factors to monomer elution from resin based dental materials. Among these factors, chemistry of the solvent, size and chemical nature of released components and extent of polymerization are the most significant ones.8 It is well defined that in order to minimize the amount of residual monomer elution, resin based restorative materials have to be polymerized to a high degree.25 Yet, it is indicated that a complete monomer conversion is never succeeded and the degree of conversion varies between 35% and 77%.8,26 Consequently, curing source intensity, wavelength, exposure duration, thickness of overlying material tip to tooth curing distance has been a topic of investigations to determine the optimal conditions for resin polymerization.2 It is widely accepted that, an intensity of 300 mW/cm2 or greater at a wavelength range of 450–500 nm is required for complete polymerization of a 2 mm composite resin,27 although different intensities and exposure time are suggested by other researchers.28,29 Since the introduction of LED devices in restorative dentistry, there has been an increased interest in comparing their ability with regular halogen lights.30 This study used a standard halogen light intensity of 400 mW/cm2. A2 colorshade was chosen for the composite resins to minimize the effect of colorants on light penetration.31 In the present study two composite resins and two fissure sealants were studied after polymerization with different curing units and exposure time at 1, 3 and 7 days time intervals. There was no BisGMA in Enamel Loc and there was no TEGDMA in Filtek Z250’s chemical compositions based on the information received from the manufacturers. However, all samples showed detectable levels of BisGMA and TEGDMA except for Filtek Z250 which released very small amount of TEGDMA with almost the same value of standard deviations that remained same for all time intervals. This result is line with manufacturer information regarding that Filtek Z250 does not contain TEGDMA as base monomers. Bis-GMA is an ingredient in many dental resins and pit and fissure sealants. In Enamel Loc group, detection of BisGMA was possible although it was not stated in the chemical formulation. By the time we started our study Enamel Loc was available at the market and was known as a relatively new self etching sealant. However, at the time of the completion of the study, it was discontinued by the manufacturer. It was claimed that heat and cold temperature extremes may cause phase separation in the resin. The detection of BisGMA could be attributed to changes in the chemical composition of the material and high sensitivity of HPLC analysis as well. Pit and fissure sealant materials contain more resin matrix and less filler particles than other dental composite materials which results in containing more monomer. HPLC analysis is one of the most trustable and commonly used separation method among other techniques.32 In the present study HPLC analysis was used to evaluate the monomer release from 2 different composite resins and fissure sealants. Elution of residual monomers from dental materials depends on the chemical composition of leachable substances as well as the chemistry of the solvent.8 Tanaka et al14 revealed that, small molecular weight monomers such as TEGDMA could be eluted more than the higher weight molecules, namely BisGMA and UDMA for their faster mobility. In the present study, the amount of TEGDMA was found to be higher than the amount of BisGMA for all dental materials. This could derive from differences in chemical properties and reactive potentials of TEGDMA and BisGMA to different curing methods.33 The elution values reported in this study are higher than previous studies.8,23 This is because the specimens were placed into the solvent right after polymerization. This is clinically relevant and corresponds to a period when further curing of material is occurring, which would tend to bind more of the unreacted molecules. It is also indicated that the rate and extent of elution of monomer is greater in organic solvents compared to pure water. However it should be noted that, unreacted monomer leaching in an organic solvent does not imitate the clinical conditions yet presents the potentially leachable portions of dental materials.34 In the present study ethanol was used as an organic solvent to measure the elution of BisGMA and TEGDMA from composite resins and fissure sealants sample. Acetonitrile was used as the solvent and mobile phase since BisGMA and TEGDMA are hydrophobic and it has been used in previous studies concerning HPLC analysis of resin restorative materials.35 Manufacturers’ recommendation for curing time for Helioseal F, Enamel Loc and Filtek Z250 is 20 s whereas for Leaddent it is 40 s. Based on the differences in polymerization conditions in dental practices, we used different curing units and exposure time. In the present study, 20 and 40 s for LED unit and 40 s curing time for halogen light was used for the 1 mm thickness of tested resin based materials. For the sealant groups, polymerization with LED for 40 s yielded less monomer elution than the recommended 20 s. For the composite resin groups, recommended 20 s curing time for Filtek Z250 resulted with same amount of elution compared to 40 s of curing with LED and halogen light. However, 40 s of curing with LED eluted less in comparison to halogen 40 s.

CONCLUSIONS

Monomer release from resin based dental materials is detectable and curing performance of different curing units and exposure time can be detected by HPLC in time related elution of monomers. Therefore, efficiency of the curing unit and applying the recommended curing time of the light activated resin based dental materials is very important to protect the patient from potential hazards of residual monomers.
Table 4

Mean TEGDMA and BisGMA release (concg/L) from Enamel Loc fissure sealant samples in to ethanol after 1,3 and 7 days. a,b,c,d,e,f,g,h P<.05.

Immersion time (days)TEGDMA (concg/l)BisGMA (concg/l)
LED 20 mean±sd (n=3)LED 40 mean±sd (n=3)QTH 40 mean±sd (n=3)LED 20 mean±sd (n=3)LED 40 mean±sd (n=3)QTH 40 mean±sd (n=3)
10.26a±0.030.11a±0.040.19±0.010.27b,c±0.080.09b±0.010.06c±0.01
30.36d±0.050.18d±0.040.25±0.020.46f,e±0.170.15f±0.020.1e±0.01
70.43±0.030.42±0.080.39±0.030.55g,h±0.110.33g±0.060.17h±0.03
Table 5

Mean TEGDMA and BisGMA release (concg/L) from Filtek Z 250 composite resin samples in to ethanol after 1,3 and 7 days. a,b P<.05.

Immersion time (days)TEGDMA (concg/l)BisGMA (concg/l)
LED 20 mean±sd (n=3)LED 40 mean±sd (n=3)QTH 40 mean±sd (n=3)LED 20 mean±sd (n=3)LED 40 mean±sd (n=3)QTH 40 mean±sd (n=3)
10.01±0.000.01±0.00.01±0.010.04±0.010.03±0.010.05±0.02
30.01±0.010.01±0.010.01±0.010.06±0.010.04a±0.010.07a±0.01
70.01±0.010.01±0.010.01±0.010.07±0.000.05b±0.010.07b±0.01
  34 in total

1.  Light-curing technology: past, present, and future.

Authors:  Paul D Hammesfahr; Michael T O'Connor; Xiuling Wang
Journal:  Compend Contin Educ Dent       Date:  2002-09

2.  Composite conversion and temperature rise using a conventional, plasma arc, and an experimental blue LED curing unit.

Authors:  Z Tarle; A Meniga; A Knezević; J Sutalo; M Ristić; G Pichler
Journal:  J Oral Rehabil       Date:  2002-07       Impact factor: 3.837

3.  Minimizing patients' exposure to uncured components in a dental sealant.

Authors:  F A Rueggeberg; M Dlugokinski; J W Ergle
Journal:  J Am Dent Assoc       Date:  1999-12       Impact factor: 3.634

4.  Time-based elution of TEGDMA and BisGMA from resin composite cured with LED, QTH and high-intensity QTH lights.

Authors:  Adil Nalçaci; Nuran Ulusoy; Orhan Atakol
Journal:  Oper Dent       Date:  2006 Mar-Apr       Impact factor: 2.440

5.  Output from visible-light activation units and depth of cure of light-activated composites.

Authors:  J F McCabe; T E Carrick
Journal:  J Dent Res       Date:  1989-11       Impact factor: 6.116

Review 6.  Clinical guidelines for photocuring restorative resins.

Authors:  W F Caughman; F A Rueggeberg; J W Curtis
Journal:  J Am Dent Assoc       Date:  1995-09       Impact factor: 3.634

Review 7.  Effects of dental restorative materials: local and systemic responses reviewed.

Authors:  H R Stanley
Journal:  J Am Dent Assoc       Date:  1993-10       Impact factor: 3.634

8.  Effect of light intensity and exposure duration on cure of resin composite.

Authors:  F A Rueggeberg; W F Caughman; J W Curtis
Journal:  Oper Dent       Date:  1994 Jan-Feb       Impact factor: 2.440

9.  Dental composite depth of cure with halogen and blue light emitting diode technology.

Authors:  R W Mills; K D Jandt; S H Ashworth
Journal:  Br Dent J       Date:  1999-04-24       Impact factor: 1.626

10.  Cytotoxic effects of dental composites, adhesive substances, compomers and cements.

Authors:  A Schedle; A Franz; X Rausch-Fan; A Spittler; T Lucas; P Samorapoompichit; W Sperr; G Boltz-Nitulescu
Journal:  Dent Mater       Date:  1998-11       Impact factor: 5.304

View more
  10 in total

1.  Cytotoxicity of resin composites containing bioactive glass fillers.

Authors:  Satin Salehi; Fernanda Gwinner; John C Mitchell; Carmem Pfeifer; Jack L Ferracane
Journal:  Dent Mater       Date:  2015-01-03       Impact factor: 5.304

2.  Evaluation of residual monomer release after polymerization of different restorative materials used in pediatric dentistry.

Authors:  Gülsüm Duruk; Sibel Akküç; Yılmaz Uğur
Journal:  BMC Oral Health       Date:  2022-06-13       Impact factor: 3.747

3.  Comparison of the Degree of Conversion of Resin Based Endodontic Sealers Using the DSC Technique.

Authors:  Elisabetta Cotti; Paola Scungio; Claudia Dettori; Guido Ennas
Journal:  Eur J Dent       Date:  2011-04

4.  Hardness, polymerization depth, and internal adaptation of Class II silorane composite restorations as a function of polymerization protocol.

Authors:  Janaina Bechtold; Priscila Jaques Dos Santos; Andrea Anido-Anido; Vinícius Di Hipólito; Roberta Caroline Bruschi Alonso; Paulo Henrique Perlatti D'Alpino
Journal:  Eur J Dent       Date:  2012-04

5.  Investigation of eluted monomers from resin-based root canal sealer by high-performance liquid chromatography analysis.

Authors:  Huma Omurlu; Hacer Deniz Arisu; Evrim Eliguzeloglu Dalkilic; Ugur Tamer; Hilal Torul
Journal:  Eur J Dent       Date:  2016 Jan-Mar

6.  Degree of Conversion and Oxygen-Inhibited Layer Effect of Three Dental Adhesives.

Authors:  Lindsay Robertson; Melissa Phaneuf; Asmaa Haimeur; Igor Pesun; Rodrigo França
Journal:  Dent J (Basel)       Date:  2016-10-27

7.  Assessment of Cytotoxic and Genotoxic Effect of Fissure Sealants in Buccal Epithelial Cells.

Authors:  Lidia Gavić; Kristina Goršeta; Anita Buterin; Domagoj Glavina; Davor Želježić; Antonija Tadin
Journal:  Acta Stomatol Croat       Date:  2021-03

8.  Evaluation of degree of conversion and hardness of dental composites photo-activated with different light guide tips.

Authors:  Marília Regalado Galvão; Sergei Godeiro Fernandes Rabelo Caldas; Vanderlei Salvador Bagnato; Alessandra Nara de Souza Rastelli; Marcelo Ferrarezi de Andrade
Journal:  Eur J Dent       Date:  2013-01

9.  The Influence of Irradiation Time and Layer Thickness on Elution of Triethylene Glycol Dimethacrylate from SDR® Bulk-Fill Composite.

Authors:  Ryta Łagocka; Katarzyna Jakubowska; Dariusz Chlubek; Jadwiga Buczkowska-Radlińska
Journal:  Biomed Res Int       Date:  2016-06-06       Impact factor: 3.411

10.  Leaching of monomers from bulk-fill composites: An in vitro study.

Authors:  Ankit Rajesh Sajnani; Mithra Nidharsh Hegde
Journal:  J Conserv Dent       Date:  2016 Sep-Oct
  10 in total