Literature DB >> 20042350

Preoperative and intraoperative sonographic visibility of collagen-based breast biopsy marker clips.

Peter R Eby1, Kristine E Calhoun, Brenda F Kurland, Wendy B Demartini, Robert L Gutierrez, Sue Peacock, Benjamin O Anderson, David R Byrd, Gary N Mann, Constance D Lehman.   

Abstract

RATIONALE AND
OBJECTIVES: The aim of this study was to determine the sonographic visibility of implanted collagen-based breast biopsy marker clips in the clinic and operating room.
MATERIALS AND METHODS: Female patients aged > or =18 years who presented for preoperative surgical evaluation within 4 weeks of ultrasound-guided breast biopsy and collagen-based marker clip placement were eligible for this pilot study. The sonographic visibility of the marker clips was rated from 1 (not visible) to 5 (clearly visible) by surgeons at the preoperative appointment, by radiologists at wire localization, and by surgeons in the operating room. Visibility was considered inadequate for values of 1 or 2 and adequate for values of 3, 4, or 5. The data were compared using Wilcoxon's signed-rank test for paired differences across physician (radiologist vs surgeon), time (preoperative visit vs day of surgery), and target (lesion vs clip).
RESULTS: Twenty-five patients with 26 lesions were enrolled, and 19 patients returned for all imaging procedures. The mean lesion size was 12 mm (range, 5-24 mm). Adequate marker clip visibility assessed by the surgeons decreased from 80% (20 of 25) at the preoperative appointment to 65% (11 of 17) in the operating room, but the difference was not significant (P=.27). Visibilities of the lesions and clips were similar at the preoperative appointment (P=.61), but the clips were significantly less visible on the day of operation (P=.03).
CONCLUSION: The sonographic visibility of collagen-based marker clips is variable and likely decreases over time but may be adequate to guide intraoperative surgical excision in many cases.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 20042350     DOI: 10.1016/j.acra.2009.10.017

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  3 in total

1.  Sonographic visibility of breast tissue markers: a tissue phantom comparison study.

Authors:  James Han-Su Seow; Michael Phillips; Donna Taylor
Journal:  Australas J Ultrasound Med       Date:  2015-12-31

2.  Hygroscopic sonographically detectable clips form characteristic breast and lymph node pseudocysts.

Authors:  Moshe Carmon; Sofia Zilber; David Gekhtman; Oded Olsha; Tal Hadar; Eliahu Golomb
Journal:  Mod Pathol       Date:  2017-08-11       Impact factor: 7.842

Review 3.  Preoperative localization and surgical margins in conservative breast surgery.

Authors:  F Corsi; L Sorrentino; D Bossi; A Sartani; D Foschi
Journal:  Int J Surg Oncol       Date:  2013-08-05