Lesion size inaccuracies in digital mammography.

Jean R Paquelet1, R Edward Hendrick.   

Abstract

OBJECTIVE: We show with both a clinical case and measurements using the American College of Radiology mammography phantom that some digital mammography acquisition and image display combinations lead to both marked over- and underestimation of lesion sizes on geometrically magnified images.
CONCLUSION: The results of this study indicate that the accuracy of lesion size measurements in all magnification modes should be a routine quality control acceptance test with each acquisition-display system combination in digital mammography.

Mesh:

Year:  2010        PMID: 20028882     DOI: 10.2214/AJR.09.2927

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959