Dose reduction for semi-automated volumetry of hepatic metastasis in MDCT studies.

Sebastian Keil1, Cedric Plumhans, Istvan A Nagy, Katharina Schiffl, Grzegorz Soza, Florian F Behrendt, Andreas H Mahnken, Rolf W Günther, Marco Das.   

Abstract

OBJECTIVES: To investigate the performance of semi-automated measurements (RECIST, volume) of hepatic metastases in multidetector-row computed tomography (MDCT) under normal-dose- and simulated low-dose-protocols.
MATERIALS AND METHODS: Thirty-five patients (67 +/- 13 years) with a total of 79 hepatic metastases underwent 16-MDCT (120 kv, 160 mAseff, pitch 1, 3 mm slice thickness, 2 mm reconstruction increment, B30f standard soft tissue kernel) for either initial staging or therapy monitoring. Corresponding raw data from these standard-dose scans were simulated at lower radiation doses of 80/60/40 mAseff (Somatom Noise Vers.6.1 beta, Siemens Healthcare, Forchheim, Germany). A semi-automated software tool (SyngoCT Oncology, Siemens Healthcare, Forchheim, Germany) was applied to each dose setting to evaluate size parameters (RECIST, volume). These measurements were compared by applying repeated-measures analysis of variance and displayed graphically.
RESULTS: For RECIST measurements no statistically significant differences were found between standard dose (Mean RECIST diameter: 20.46 +/- 8.37 mm) and different simulated low radiation doses (80 mAseff: 20.95 +/- 8.20 mm/60 mAseff: 20.50 +/- 8.35 mm/40 mAseff: 19.95 +/- 8.16 mm): P = 0.0774.Statistically significant differences of volume quantification (P < 0.05) could be found between standard-(3.60 +/- 4.63 mL) and simulated lowest dose of 40 mAseff (3.17 +/- 4.08 mL), whereas there was no difference (P > 0.05) between 160 mAseff- and either 80 mAseff-(3.46 +/- 4.31 mL) or 60 mAseff-protocols (3.44 +/- 4.35 mL).
CONCLUSIONS: Software-assisted assessment of RECIST criteria and volume demonstrated valid performances under different dose-settings in MDCT; therefore, substantial radiation dose reduction could be possible with the use of semi-automated measurements in follow-up studies.

Entities:  

Mesh:

Year:  2010        PMID: 20027116     DOI: 10.1097/RLI.0b013e3181cd194e

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  1 in total

1.  Towards volumetric thresholds in RECIST 1.1: Therapeutic response assessment in hepatic metastases.

Authors:  Katharina S Winter; Felix O Hofmann; Kolja M Thierfelder; Julian W Holch; Nina Hesse; Alena B Baumann; Dominik P Modest; Sebastian Stintzing; Volker Heinemann; Jens Ricke; Wieland H Sommer; Melvin D'Anastasi
Journal:  Eur Radiol       Date:  2018-05-07       Impact factor: 5.315

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.