Literature DB >> 20025045

Drug-eluting stent fracture: incidence, contributing factors, and clinical implications.

Timothy Canan1, Michael S Lee.   

Abstract

Stent fracture has been observed in noncoronary vessels, especially in the superficial femoral and popliteal arteries and with bare metal stents in saphenous vein grafts of coronary arteries. Since the introduction of drug-eluting stents, stent fractures have also been reported in small studies and case reports. We reviewed these publications to assess what is known regarding the incidence, contributing factors, and clinical implications of drug-eluting stent fracture in coronary arteries. The reported rate of drug-eluting stent fracture in coronary arteries ranges from 1 to 8%, although much of the available literature is derived from single-center studies that are heterogeneous in their study methods. A higher risk of stent fracture may be associated with the right coronary artery location, excessive tortuosity or angulation of the vessel, overlapping stents, and longer stents. The closed-cell design of the Cypher stent has been associated with increased rigidity that may increase the risk of stent fracture, although these studies did not assess the overall outcomes between the Cypher and Taxus stents in a head-to-head comparison. Stent fracture has been shown by most studies to be associated with a statistically increased incidence of focal in-stent restenosis, and some have shown an increased risk of target lesion revascularization. Other complications observed with stent fracture include stent thrombosis, coronary aneurysms, myocardial infarction, and sudden death.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20025045     DOI: 10.1002/ccd.22212

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  14 in total

1.  Circumferential stent fracture: novel detection and treatment with the use of StentBoost.

Authors:  Raghu T Ramegowda; Srinivas B Chikkaswamy; Ashalatha Bharatha; Jayashree Radhakrishna; Geetha B Krishnanaik; Manjunath C Nanjappa; Arunkumar Panneerselvam
Journal:  Tex Heart Inst J       Date:  2012

Review 2.  Are we aware of stent fracture?

Authors:  R D Acar; M Bulut; M Akcakoyun
Journal:  Herz       Date:  2013-10-25       Impact factor: 1.443

3.  Intravascular Ultrasound-Guided Treatment for In-stent Restenosis Associated with Stent Fracture in Overlapped Drug-eluting Stents.

Authors:  Min Chul Kim; Myung Ho Jeong; Doo Sun Sim; Young Joon Hong; Ju Han Kim; Youngkeun Ahn
Journal:  Chonnam Med J       Date:  2019-09-24

4.  Stent fracture after everolimus-eluting stent implantation.

Authors:  Ali S Almasood; Xavier Freixa; Sohail Q Khan; Peter H Seidelin; Vladimír Džavík
Journal:  Cardiol Res Pract       Date:  2011-04-03       Impact factor: 1.866

5.  Clinical characteristics of coronary drug-eluting stent fracture: insights from a two-center des registry.

Authors:  Kyung Woo Park; Jin Joo Park; In-Ho Chae; Jae-Bin Seo; Han-Mo Yang; Hae-Young Lee; Hyun-Jae Kang; Young-Seok Cho; Tae-Jin Yeon; Woo-Young Chung; Bon-Kwon Koo; Dong-Ju Choi; Byung-Hee Oh; Young-Bae Park; Hyo-Soo Kim
Journal:  J Korean Med Sci       Date:  2010-12-22       Impact factor: 2.153

6.  Successful management of a rare case of stent fracture and subsequent migration of the fractured stent segment into the ascending aorta in in-stent restenotic lesions of a saphenous vein graft.

Authors:  Hoyoun Won; Jaewon Oh; Youngjun Yang; Mihyun Kim; Choongki Kim; Junbeom Park; Byeong-Keuk Kim; Donghoon Choi; Myeong-Ki Hong
Journal:  Korean Circ J       Date:  2012-01-31       Impact factor: 3.243

7.  Complete stent fracture 1 year after LIMA PCI due to LIMA and subclavian artery dissection.

Authors:  Dimitris Anagnostou; Elias Sanidas; Ioannis Paizis; John Barbetseas
Journal:  Oxf Med Case Reports       Date:  2015-09-01

8.  Zotarolimus-eluting stent fracture at initial implantation diagnosed with StentBoost.

Authors:  Alev Arat Ozkan; Umit Yasar Sinan; Aziz T Gurmen
Journal:  SAGE Open Med Case Rep       Date:  2016-05-11

9.  Stent fracture: how frequently is it recognized?

Authors:  Mohammed Khalil Mohsen; Awad Alqahtani; Jassim Al Suwaidi
Journal:  Heart Views       Date:  2013-04

Review 10.  Coronary stent fracture: a recently appreciated phenomenon with clinical relevance.

Authors:  Madjid Chinikar; Parham Sadeghipour
Journal:  Curr Cardiol Rev       Date:  2014-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.