Literature DB >> 20013279

What should we do with a discontinued shunt?

Yun-Ho Lee1, Eun Kyung Park, Dong-Seok Kim, Joong-Uhn Choi, Kyu-Won Shim.   

Abstract

PURPOSE: The reported rate is up to 10% of shunt disconnection or fracture, either ventriculoperitoneal or subduroperitoneal. However, not all of shunt discontinuity is associated with shunt malfunction. We analyzed the discontinuity of the shunt system and related factors and tried to present a follow-up policy.
METHODS: This is a retrospective study involving 22 patients who presented with shunt disconnection because of malfunction or other reasons between January 2003 and October 2008. To evaluate shunt function, we performed a shuntogram, temporary ligation, or intraoperative exploration.
RESULTS: Nine (40.9%) of 22 patients had nonfunctioning shunts. The shunt system was removed in eight cases, and one patient refused surgical management. Of these 22 patients with disconnected shunts, 17 shunts placed in the occipital area were discontinued. The discontinuations were mostly fractured at neck (52%), not disconnected at connection site.
CONCLUSION: Shunt disconnection has not been always represented a shunt malfunction. We thought many patients with disconnected shunt may be independent from shunt system, and it can be a good chance for patient to remove the shunt system.

Entities:  

Mesh:

Year:  2009        PMID: 20013279     DOI: 10.1007/s00381-009-1061-x

Source DB:  PubMed          Journal:  Childs Nerv Syst        ISSN: 0256-7040            Impact factor:   1.475


  17 in total

1.  Follow-up of children with shunted hydrocephalus.

Authors:  A Colak; A L Albright; I F Pollack
Journal:  Pediatr Neurosurg       Date:  1997-10       Impact factor: 1.162

2.  [Arrested shunt-independent hydrocephalus].

Authors:  G Kaiser
Journal:  Z Kinderchir       Date:  1983-04

3.  Factors causing acute shunt infection. Computer analysis of 1174 operations.

Authors:  D Renier; J Lacombe; A Pierre-Kahn; C Sainte-Rose; J F Hirsch
Journal:  J Neurosurg       Date:  1984-12       Impact factor: 5.115

4.  Danish experience with the one-piece shunt. A long-term follow-up.

Authors:  J Haase; F Bang; M Tange
Journal:  Childs Nerv Syst       Date:  1987       Impact factor: 1.475

5.  Complications of ventriculo-peritoneal shunting and a critical comparison of the three-piece and one-piece systems.

Authors:  A J Raimondi; J S Robinson; K Kuwawura
Journal:  Childs Brain       Date:  1977

6.  Cerebrospinal fluid shunt infection. Influences on initial management and subsequent outcome.

Authors:  B C Walters; H J Hoffman; E B Hendrick; R P Humphreys
Journal:  J Neurosurg       Date:  1984-05       Impact factor: 5.115

7.  Randomized trial of cerebrospinal fluid shunt valve design in pediatric hydrocephalus.

Authors:  J M Drake; J R Kestle; R Milner; G Cinalli; F Boop; J Piatt; S Haines; S J Schiff; D D Cochrane; P Steinbok; N MacNeil
Journal:  Neurosurgery       Date:  1998-08       Impact factor: 4.654

8.  Time-related patterns of ventricular shunt failure.

Authors:  J Kast; D Duong; F Nowzari; W M Chadduck; S J Schiff
Journal:  Childs Nerv Syst       Date:  1994-11       Impact factor: 1.475

View more
  2 in total

1.  Disconnected subduroperitoneal shunt catheter induces silent bowel perforation: An unusual complication.

Authors:  Hsiang-Ming Huang; Wen-Yuan Lee; Der-Cherng Chen
Journal:  Int J Surg Case Rep       Date:  2011-03-02

2.  The disconnected shunt: a window of opportunities.

Authors:  Jonathan Roth; Shlomi Constantini
Journal:  Childs Nerv Syst       Date:  2017-02-24       Impact factor: 1.475

  2 in total