Fluid dynamics of cytotoxic safety cabinets.

R Braconnier1, F Bonthoux.   

Abstract

This study investigated the specific fluid dynamics characteristics of cytotoxic safety cabinets (CSC), particularly those used in cancer drug reconstitution operations. Measurements taken on site were used to derive characteristic data for these cabinets. An in-depth laboratory investigation of airflows inside another CSC was also conducted. Anemometric values recorded on these two installations enabled the experimental validation of computational fluid dynamics methods applied to CSC. The digital flow simulations conducted provide a better understanding of the detailed flow structure inside a CSC and made it possible to study the influence of different operating parameters on the air velocity distribution inside the cabinet front opening: recycled air temperature, product protection airflow rate, suction openings spatial distribution, air compensation mode and draughts, operator arm penetration, and operator presence in front of the cabinet.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 20007340     DOI: 10.1093/annhyg/mep087

Source DB:  PubMed          Journal:  Ann Occup Hyg        ISSN: 0003-4878


  1 in total

1.  Analysis of factors affecting containment with extracted partial enclosures using computational fluid dynamics.

Authors:  Rachel L Batt; Adrian Kelsey
Journal:  Ann Occup Hyg       Date:  2013-11-14