Literature DB >> 19997298

A terahertz time-domain spectrometer for simultaneous transmission and reflection measurements at normal incidence.

Fabian D J Brunner1, Arno Schneider, Peter Günter.   

Abstract

We present a versatile terahertz time-domain spectrometer which allows reflection measurements at normal incidence and double pass transmission measurements in a single experimental setup. Two different modes for transmission measurements are demonstrated for precise measurements of transparent high or low refractive index materials, respectively. The refractive indices and absorption coefficients of cesium iodide, potassium bromide, sodium chloride, polytetrafluoroethylene (PTFE, Teflon), and silicon have been measured in the frequency range between 1.4 and 4.7 THz. The parameters of the Lorentz oscillator functions describing the phonon polariton dispersions of CsI and KBr have been determined.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19997298     DOI: 10.1364/OE.17.020684

Source DB:  PubMed          Journal:  Opt Express        ISSN: 1094-4087            Impact factor:   3.894


  1 in total

1.  Co-registered combined OCT and THz imaging to extract depth and refractive index of a tissue-equivalent test object.

Authors:  A J Fitzgerald; X Tie; M J Hackmann; B Cense; A P Gibson; V P Wallace
Journal:  Biomed Opt Express       Date:  2020-02-19       Impact factor: 3.732

  1 in total