Fast Monte Carlo simulation on a voxelized human phantom deformed to a patient.

G Bueno1, O Déniz, C B Carrascosa, J M Delgado, L Brualla.   

Abstract

PURPOSE: A method for performing fast simulations of absorbed dose using a patient's computerized tomography (CT) scan without explicitly relying on a calibration curve is presented.
METHODS: The method is based on geometrical deformations performed on a standard voxelized human phantom. This involves spatially transforming the human phantom to align it with the patient CT image. Since the chemical composition and density of each voxel are given in the phantom data, a calibration curve is not used in the proposed method. For this study, the Monte Carlo (MC) code PENELOPE has been used as the simulation of reference. The results obtained with PENELOPE simulations are compared to those obtained with PENFAST and with the collapsed cone convolution algorithm implemented in a commercial treatment planning system.
RESULTS: The comparisons of the absorbed doses calculated with the different algorithms on two patient CTs and the corresponding deformed phantoms show a maximum distance to agreement of 2 mm, and in general, the obtained absorbed dose distributions are compatible within the reached statistical uncertainty. The validity of the deformation method for a broad range of patients is shown using MC simulations in random density phantoms. A PENFAST simulation of a 6 MV photon beam impinging on a patient CT reaches 2% statistical uncertainty in the absorbed dose, in a 0.1 cm3 voxel along the central axis, in 10 min running on a single core of a 2.8 GHz CPU.
CONCLUSIONS: The proposed method of the absorbed dose calculation in a deformed voxelized phantom allows for dosimetric studies in the geometry of a patient CT scan. This is due to the fact that the chemical composition and material density of the phantom are known. Furthermore, simulation using the phantom geometry can provide dosimetric information for each organ. The method can be used for quality assurance procedures. In relation to PENFAST, it is shown that a purely condensed-history algorithm (class I) can be used for absorbed dose estimation in patient CTs.

Entities:  

Mesh:

Year:  2009        PMID: 19994526      PMCID: PMC2774349          DOI: 10.1118/1.3245877

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  19 in total

1.  Correlation between CT numbers and tissue parameters needed for Monte Carlo simulations of clinical dose distributions.

Authors:  W Schneider; T Bortfeld; W Schlegel
Journal:  Phys Med Biol       Date:  2000-02       Impact factor: 3.609

2.  A revision of the gamma-evaluation concept for the comparison of dose distributions.

Authors:  Annemarie Bakai; Markus Alber; Fridtjof Nüsslin
Journal:  Phys Med Biol       Date:  2003-11-07       Impact factor: 3.609

3.  All about MAX: a male adult voxel phantom for Monte Carlo calculations in radiation protection dosimetry.

Authors:  R Kramer; J W Vieira; H J Khoury; F R A Lima; D Fuelle
Journal:  Phys Med Biol       Date:  2003-05-21       Impact factor: 3.609

4.  Effects of density changes in the chest on lung stereotactic radiotherapy.

Authors:  Tatsuya Fujisaki; Kyoko Kikuchi; Hidetoshi Saitoh; Naoki Tohyama; Atsushi Myojoyama; Atsushi Osawa; Akio Kuramoto; Shinji Abe; Tesuo Inada; Takatsugu Kawase; Etsuo Kunieda
Journal:  Radiat Med       Date:  2004 Jul-Aug

5.  Efficient Monte Carlo simulation of multileaf collimators using geometry-related variance-reduction techniques.

Authors:  L Brualla; F Salvat; R Palanco-Zamora
Journal:  Phys Med Biol       Date:  2009-06-12       Impact factor: 3.609

6.  The calibration of CT Hounsfield units for radiotherapy treatment planning.

Authors:  U Schneider; E Pedroni; A Lomax
Journal:  Phys Med Biol       Date:  1996-01       Impact factor: 3.609

7.  A technique for the quantitative evaluation of dose distributions.

Authors:  D A Low; W B Harms; S Mutic; J A Purdy
Journal:  Med Phys       Date:  1998-05       Impact factor: 4.071

8.  Comparison of two methods for assessing patient dose from computed tomography.

Authors:  J Geleijns; J G Van Unnik; J Zoetelief; D Zweers; J J Broerse
Journal:  Br J Radiol       Date:  1994-04       Impact factor: 3.039

9.  A convolution method of calculating dose for 15-MV x rays.

Authors:  T R Mackie; J W Scrimger; J J Battista
Journal:  Med Phys       Date:  1985 Mar-Apr       Impact factor: 4.071

10.  The implementation of a generalised Batho inhomogeneity correction for radiotherapy planning with direct use of CT numbers.

Authors:  K J Cassell; P A Hobday; R P Parker
Journal:  Phys Med Biol       Date:  1981-09       Impact factor: 3.609

View more
  4 in total

1.  A geodesic deformable model for automatic segmentation of image sequences applied to radiation therapy.

Authors:  G Bueno; O Déniz; J Salido; C Carrascosa; J M Delgado
Journal:  Int J Comput Assist Radiol Surg       Date:  2010-07-20       Impact factor: 2.924

2.  PRIMO: a graphical environment for the Monte Carlo simulation of Varian and Elekta linacs.

Authors:  M Rodriguez; J Sempau; L Brualla
Journal:  Strahlenther Onkol       Date:  2013-09-06       Impact factor: 3.621

Review 3.  Monte Carlo systems used for treatment planning and dose verification.

Authors:  Lorenzo Brualla; Miguel Rodriguez; Antonio M Lallena
Journal:  Strahlenther Onkol       Date:  2016-11-25       Impact factor: 3.621

4.  A high-resolution dose calculation engine for X-ray microbeams radiation therapy.

Authors:  Sarvenaz Keshmiri; Sylvan Brocard; Raphaël Serduc; Jean-François Adam
Journal:  Med Phys       Date:  2022-04-12       Impact factor: 4.506

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.