Literature DB >> 19964867

Evaluation of a robotic arm for echocardiography to X-ray image registration during cardiac catheterization procedures.

Yingliang Ma1, Graeme P Penney, Dennis Bos, Peter Frissen, George de Fockert, Andy King, Gang Gao, Cheng Yao, John Totman, Matthew Ginks, C Rinaldi, Reza Razavi, Kawal S Rhode.   

Abstract

We present an initial evaluation of a robotic arm for positioning a 3D echo probe during cardiac catheterization procedures. By tracking the robotic arm, X-ray table and X-ray C-arm, we are able to register the 3D echo images with live 2D X-ray images. In addition, we can also use tracking data from the robotic arm combined with system calibrations to create extended field of view 3D echo images. Both these features can be used for roadmapping to guide cardiac catheterization procedures. We have carried out a validation experiment of our registration method using a cross-wire phantom. Results show our method to be accurate to 3.5 mm. We have successfully demonstrated the creation of the extended field of view data on 2 healthy volunteers and the registration of echo and X-ray data on 1 patient undergoing a pacing study.

Entities:  

Mesh:

Year:  2009        PMID: 19964867     DOI: 10.1109/IEMBS.2009.5334392

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X


  1 in total

1.  Analysis of a Customized Clutch Joint Designed for the Safety Management of an Ultrasound Robot.

Authors:  Shuangyi Wang; Richard James Housden; Yohan Noh; Anisha Singh; Lukas Lindenroth; Hongbin Liu; Kaspar Althoefer; Joseph Hajnal; Davinder Singh; Kawal Rhode
Journal:  Appl Sci (Basel)       Date:  2019-05-09       Impact factor: 2.679