Literature DB >> 19964245

Laboratory prototype of cochlear implant: design and techniques.

Hussnain Ali1, Talha J Ahmad, Asim Ajaz, Shoab A Khan.   

Abstract

This paper presents design overview of a low cost prototype of Cochlear Implant developed from commercial off-the-shelf components. Design scope includes speech processing module implemented on a commercial digital signal processor, transcutaneous data and power transceiver developed from a single pair of inductive coils and finally a stimulator circuitry for cochlear stimulation. Different speech processing strategies such as CIS, SMSP and F0/F1 have been implemented and tested using a novel, indigenously developed speech processing research module which evaluates the performance of speech processing strategies in software, hardware and practical scenarios. Design overview, simulations and practical results of an optimized inductive link using Class E Power Amplifier are presented. Link was designed at a carrier frequency of 2.5MHz for 100mW output power. Receiver logic design and stimulator circuitry was implemented using a PIC microcontroller and off-the-shelf electronic components. Results indicate 40% link efficiency with 128kbps data transfer rate. This low cost prototype can be used for undertaking cochlear implant research in laboratories.

Mesh:

Year:  2009        PMID: 19964245     DOI: 10.1109/IEMBS.2009.5333707

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X


  1 in total

1.  Simplified design equations for Class-E neural prosthesis transmitters.

Authors:  Philip Troyk; Zhe Hu
Journal:  IEEE Trans Biomed Eng       Date:  2013-01-01       Impact factor: 4.538

  1 in total