| Literature DB >> 19963977 |
Nathalia Peixoto1, Kassandra Jackson, Raamin Samiyi, Saugandhika Minnikanti.
Abstract
Here we report on long-term (300 to 600 hours) stability measures for implantable stimulating electrodes. We have considered several measures of stability as they refer to reliability of charge carrying capacity in implantable electrodes. We have designed and manufactured coatings for large area (1 to 2mm(2)) stainless steel substrates. Materials tested were electrodeposited iridium oxide films, multi-walled carbon nanotube mesh, and PEDOT:PSS. Traditional characterization techniques such as cyclic voltammetry and electrochemical impedance spectroscopy cover a small fraction of the characterization framework needed for ensuring the safety and performance of electrodes designed for long-term implants. The stability measures suggested here rely on continuous low frequency cycling and evaluation of cathodic charge storage capacity during cycling. We experimentally show, in this paper, that the stability may be measured and is relevant for long-term applications of such coatings.Entities:
Mesh:
Year: 2009 PMID: 19963977 DOI: 10.1109/IEMBS.2009.5333449
Source DB: PubMed Journal: Conf Proc IEEE Eng Med Biol Soc ISSN: 1557-170X