Literature DB >> 19961633

In vitro delivery of budesonide from 30 jet nebulizer/compressor combinations using infant and child breathing patterns.

Elna B Berg1, Robert J Picard.   

Abstract

BACKGROUND: An aerosol of budesonide inhalation suspension is delivered when used with various jet-nebulizer/compressor combinations. The constant introduction of new nebulizer/compressor combinations raises the question of whether the performance of these match the performance of existing devices. The aim of this study was to determine in vitro the inhaled mass and aerosol characteristics of budesonide inhalation suspension from a selection of jet-nebulizer/compressor combinations presently marketed in the United States, Europe, and Japan.
METHODS: The in vitro characterization was performed using standardized and published methods. Each nebulizer was charged with 1 vial (2 mL) of budesonide inhalation suspension 0.25 mg/mL (0.5 mg budesonide) and run until end of aerosol formation. Droplet size and distribution was determined using a cooled impactor at air flow of 15 L/min. The inhaled mass of budesonide (ie, mass on the inhalation filter) was collected using a breathing simulator that mimicked the breathing patterns of an infant and a child. The aerosol was collected on filters placed between the nebulizer mouthpiece and the breathing simulator. Budesonide was quantified via standard high-performance liquid chromatography.
RESULTS: The mass median aerodynamic diameter of the aerosol measured with the cooled impactor ranged between 4.8 mum and 9.9 mum, and the geometric standard deviation ranged between 1.7 mum and 2.1 mum. The inhaled mass of budesonide expressed as a percentage of the nebulizer charge ranged from 1% to 9% (infant) and from 4% to 20% (child).
CONCLUSIONS: The in vitro budesonide mass collected on the inhalation filter and delivery characteristics differed considerably between the 30 nebulizer/compressor combinations. The present in vitro characterization of jet nebulizers can be used as a guidance for selection of jet-nebulizer/compressor combinations for delivery of the budesonide nebulization suspension in the home-care setting. Further investigations of new nebulizer/compressor combinations are warranted.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19961633

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  5 in total

Review 1.  A path to successful patient outcomes through aerosol drug delivery to children: a narrative review.

Authors:  Arzu Ari
Journal:  Ann Transl Med       Date:  2021-04

2.  Evaluation of a nanoemulsion-based formulation for respiratory delivery of budesonide by nebulizers.

Authors:  Amir Amani; Peter York; Henry Chrystyn; Brian J Clark
Journal:  AAPS PharmSciTech       Date:  2010-07-24       Impact factor: 3.246

3.  Variability in delivered dose and respirable delivered dose from nebulizers: are current regulatory testing guidelines sufficient to produce meaningful information?

Authors:  Ross Hm Hatley; Sarah M Byrne
Journal:  Med Devices (Auckl)       Date:  2017-02-01

4.  Effectiveness of HEPA Filters at Removing Infectious SARS-CoV-2 from the Air.

Authors:  Hiroshi Ueki; Michiko Ujie; Yosuke Komori; Tatsuo Kato; Masaki Imai; Yoshihiro Kawaoka
Journal:  mSphere       Date:  2022-08-10       Impact factor: 5.029

5.  Effectiveness of Face Masks in Preventing Airborne Transmission of SARS-CoV-2.

Authors:  Hiroshi Ueki; Yuri Furusawa; Kiyoko Iwatsuki-Horimoto; Masaki Imai; Hiroki Kabata; Hidekazu Nishimura; Yoshihiro Kawaoka
Journal:  mSphere       Date:  2020-10-21       Impact factor: 4.389

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.