Literature DB >> 19958376

Bone-supportive behavior of microplasma-sprayed CaP-coated implants: mechanical and histological outcome in the goat.

R Junker1, P J D Manders, J Wolke, Y Borisov, J A Jansen.   

Abstract

BACKGROUND: New microplasma spray (MPS) equipment to deposit calcium phosphate (CaP) ceramic coatings onto titanium substrates has been developed. With this apparatus, it is possible to spray fine particles, as well as to apply textured hydroxylapatite coatings onto titanium surfaces. Moreover, due to the low heat power of the microplasma jet, overheating of the powder particles as well as excessive local overheating of the substrate is reduced. Furthermore, because of the small laminar plasma jet, it is possible to achieve high spray efficiency in the case of spraying of dental implants. Additionally, the low level of noise (25-50 dB) and hardly any dust makes it possible to operate the MSE under normal workroom conditions.
OBJECTIVE: The aim of this study was to examine the biological properties of different MPS - CaP coatings on titanium implants when inserted into the femoral condyle of goats.
MATERIAL AND METHODS: For histomorphometrical evaluation as well as mechanical testing, 48 screw-type titanium implants were inserted into the femoral condyles of 12 goats each. The implants were either coated with a conventionally plasma-sprayed CaP ceramic, MPS CaP ceramic, or acid-etched without an additional CaP coating. Six and 12 weeks after implantation, the animals were sacrificed and bone-to-implant contact, amount of bone, as well as mechanical bone fixation were evaluated.
RESULTS: For bone-to-implant contact no statistically significant difference was found between the different CaP coatings. However, statistically significant differences were found between non-coated, acid-etched titanium implants and CaP-coated implants after 6 and 12 weeks of healing. The bone values were not statistically significantly different between the different CaP coatings at 6 and 12 weeks. Furthermore, CaP ceramic-coated implants showed statistically significantly higher torque values compared with the non-coated implants after 6 and at 12 weeks of healing. No significant differences existed between the various types of CaP coatings.
CONCLUSION: On the basis of our observations, we conclude that conventionally plasma-sprayed CaP ceramic-coated implants, as well as MPS-coated implants have a comparable effect on adjacent bone response.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19958376     DOI: 10.1111/j.1600-0501.2009.01819.x

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  6 in total

Review 1.  A systematic review on the long-term success of calcium phosphate plasma-spray-coated dental implants.

Authors:  B A J A van Oirschot; E M Bronkhorst; J J J P van den Beucken; G J Meijer; J A Jansen; R Junker
Journal:  Odontology       Date:  2016-02-17       Impact factor: 2.634

2.  Functionalizing calcium phosphate biomaterials with antibacterial silver particles.

Authors:  Jae Sung Lee; William L Murphy
Journal:  Adv Mater       Date:  2012-11-26       Impact factor: 30.849

3.  Calcium orthophosphate coatings, films and layers.

Authors:  Sergey V Dorozhkin
Journal:  Prog Biomater       Date:  2012-09-26

4.  Comparison of Physical-chemical and Mechanical Properties of Chlorapatite and Hydroxyapatite Plasma Sprayed Coatings.

Authors:  Imane Demnati; David Grossin; Olivier Marsan; Ghislaine Bertrand; Gérard Collonges; Christèle Combes; Maria Parco; Inigo Braceras; Joel Alexis; Yannick Balcaen; Christian Rey
Journal:  Open Biomed Eng J       Date:  2015-02-27

5.  In vivo and in vitro investigations of a nanostructured coating material - a preclinical study.

Authors:  Martin Adam; Cornelia Ganz; Weiguo Xu; Hamid-Reza Sarajian; Werner Götz; Thomas Gerber
Journal:  Int J Nanomedicine       Date:  2014-02-14

6.  A systematic review on the effect of inorganic surface coatings in large animal models and meta-analysis on tricalcium phosphate and hydroxyapatite on periimplant bone formation.

Authors:  Jeanne-Marie Damerau; Susanne Bierbaum; Daniel Wiedemeier; Paula Korn; Ralf Smeets; Gregor Jenny; Johanna Nadalini; Bernd Stadlinger
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2021-07-16       Impact factor: 3.405

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.