We describe our experience using a new device that results in a bloodless field in open repair of distal radius fractures. The device, an exsanguinating tourniquet (HemaClear model/40, OHK Medical Devices, Haifa, Israel), replaces the traditional methods of limb elevation, Esmarch bandaging, pneumatic tourniquet pressurizing and the associated components. HemaClear/40 is an elastic silicon ring with a tubular elastic sleeve rolled onto it. The device has attached straps that, when pulled, unroll the sleeve, rolling the ring mesially on the limb. The pressure exerted by rolling HemaClear/40 is supra-systolic thereby exsanguinating the limb and occluding the arterial inflow. Our experience in 49 patients demonstrated quick application, superior exsanguination and that the device could be placed on the forearm instead of the upper arm. No side effects or complications were noted. In our opinion, the fact that HemaClear/40 is a sterile, single-patient device makes it superior over the traditional technology.

Entities:  

Mesh:

Year:  2009        PMID: 19956041     DOI: 10.1097/BTH.0b013e3181b56187

Source DB:  PubMed          Journal:  Tech Hand Up Extrem Surg        ISSN: 1089-3393


  3 in total

1.  The Effect of the Silicone Ring Tourniquet and Standard Pneumatic Tourniquet on the Motor Nerve Conduction, Pain and Grip Strength in Healthy Volunteers.

Authors:  Georgios I Drosos; Georgios Kiziridis; Cristina Aggelopoulou; Dimitrios Galiatsatos; George Anastassopoulos; Athanasios Ververidis; Konstantinos Kazakos
Journal:  Arch Bone Jt Surg       Date:  2016-01

2.  Comparison of local pain and tissue reaction between conventional pneumatic tourniquet and disposable silicone ring tourniquet during Total Knee Arthroplasty.

Authors:  Sanjay Bhalchandra Londhe; Ravi Vinod Shah; Shubhankar Sanjay Londhe; Pritesh Omprakash Agrawal; Nicholas A Antao; Sushil Churhe
Journal:  J Clin Orthop Trauma       Date:  2020-09-09

3.  Silicone ring tourniquet versus pneumatic cuff tourniquet in carpal tunnel release: a randomized comparative study.

Authors:  G I Drosos; A Ververidis; N I Stavropoulos; R Mavropoulos; G Tripsianis; K Kazakos
Journal:  J Orthop Traumatol       Date:  2013-01-30
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.