Physical aspects of total body irradiation of bone marrow transplant patients using 18 MV x rays.

A Niroomand-Rad1.   

Abstract

The physical aspects of a total body irradiation (TBI) treatment are described. Patients seated in a special chair with their legs bent backwards are irradiated anteriorly and posteriorly (AP/PA). The chair reduces patient movement and facilitates positioning patients during 9 fractions of TBI over a 3-day period. The dose to lower extremities are monitored and raised to the total body dose. A conventional linear accelerator in a standard size treatment room is used to deliver 18 MV x-ray beams at a dose rate of approximately 20 cGy/min at a 350 cm treatment distance. Results of dose distribution, field flatness, dose uniformity, in vivo and in vitro dosimetry, and boost irradiation techniques are described.

Entities:  

Mesh:

Year:  1991        PMID: 1995548     DOI: 10.1016/0360-3016(91)90077-h

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038