Literature DB >> 19954352

Virtual reality laparoscopic nephrectomy simulator is lacking in construct validity.

R P W F Wijn1, M C Persoon, B M A Schout, E J Martens, A J J A Scherpbier, A J M Hendrikx.   

Abstract

BACKGROUND AND
PURPOSE: Several training models have been developed to improve surgeons' operative skills as well as patient outcomes. Before implementing these models in training programs, their usefulness and accuracy need to be assessed. In this study, we examined the ability of a laparoscopic nephrectomy (LN) virtual reality (VR) simulator to distinguish between different levels of expertise (construct validity).
METHODS: Twenty-two novices (no LN experience), 32 intermediates (<10 LN procedures performed) and 10 experienced urologists (> or =10 LN procedures performed) performed the same retroperitoneal task on the LN VR simulator (Mentice, Sweden) three times, performing a practice task before and after the second time. Outcome parameters were time, blood loss, path length, and total score (combination of 62 different parameters).
RESULTS: No significant differences were found between intermediate and experienced participants. Task 3 performance showed no significant difference between any of the groups. Both intermediates and experienced participants were significantly faster than novices on the first two tasks and had a better total score. Learning curves of intermediate and experienced participants were flat after task two.
CONCLUSIONS: The LN-VR simulator did not distinguish between intermediate and experienced participants. The analysis of the learning curves suggests that the tasks measured dexterity in using the simulator rather than an actual improvement of operative skills. We conclude that the LN-VR simulator does not have sufficient construct validity and is therefore, in its present form, not suitable for implementation in a urologic training program.

Entities:  

Mesh:

Year:  2010        PMID: 19954352     DOI: 10.1089/end.2009.0219

Source DB:  PubMed          Journal:  J Endourol        ISSN: 0892-7790            Impact factor:   2.942


  7 in total

Review 1.  Procedural virtual reality simulation in minimally invasive surgery.

Authors:  Cecilie Våpenstad; Sonja N Buzink
Journal:  Surg Endosc       Date:  2012-09-07       Impact factor: 4.584

Review 2.  Simulation-based training and assessment in urological surgery.

Authors:  Abdullatif Aydin; Nicholas Raison; Muhammad Shamim Khan; Prokar Dasgupta; Kamran Ahmed
Journal:  Nat Rev Urol       Date:  2016-08-23       Impact factor: 14.432

3.  Prediction of open urinary tract in laparoscopic partial nephrectomy by virtual resection plane visualization.

Authors:  Daiki Ueno; Kazuhide Makiyama; Hiroyuki Yamanaka; Takashi Ijiri; Hideo Yokota; Yoshinobu Kubota
Journal:  BMC Urol       Date:  2014-06-13       Impact factor: 2.264

4.  Measurement of the Physical Properties during Laparoscopic Surgery Performed on Pigs by Using Forceps with Pressure Sensors.

Authors:  Hiroyuki Yamanaka; Kazuhide Makiyama; Kimito Osaka; Manabu Nagasaka; Masato Ogata; Takahiro Yamada; Yoshinobu Kubota
Journal:  Adv Urol       Date:  2015-02-17

5.  LAPKaans: Tool-Motion Tracking and Gripping Force-Sensing Modular Smart Laparoscopic Training System.

Authors:  Luis H Olivas-Alanis; Ricardo A Calzada-Briseño; Victor Segura-Ibarra; Elisa V Vázquez; Jose A Diaz-Elizondo; Eduardo Flores-Villalba; Ciro A Rodriguez
Journal:  Sensors (Basel)       Date:  2020-12-04       Impact factor: 3.576

6.  Validity assessment of the laparoscopic radical nephrectomy module of the LapVision virtual reality simulator.

Authors:  Haruka Miyata; Takashige Abe; Kiyohiko Hotta; Madoka Higuchi; Takahiro Osawa; Ryuji Matsumoto; Hiroshi Kikuchi; Yo Kurashima; Sachiyo Murai; Nobuo Shinohara
Journal:  Surg Open Sci       Date:  2019-11-01

7.  Preparation for pyeloplasty for ureteropelvic junction obstruction using a patient-specific laparoscopic simulator: a case report.

Authors:  Hiroyuki Yamanaka; Kazuhide Makiyama; Tomoyuki Tatenuma; Ryoko Sakata; Futoshi Sano; Yoshinobu Kubota
Journal:  J Med Case Rep       Date:  2012-10-05
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.