Estimation of blood haemoglobin concentration using the HemoCue during caesarean section: the effect of sampling site.

N A Richards1, H Boyce, S M Yentis.   

Abstract

BACKGROUND: Haemoglobin concentration measured using the HemoCue is accurate for capillary and venous/arterial blood, provided the recommended sampling method is strictly observed. Analysis of blood, particularly of capillary samples, using the HemoCue is useful during caesarean section. The toe might be preferred to the thumb since it is numb during neuraxial anaesthesia, but whether sampling at either site is accurate in this situation, given the cardiovascular effects of anaesthesia and pregnancy, is not known. We aimed to compare haemoglobin values measured in venous and capillary samples (toe and thumb) during caesarean section under neuraxial anaesthesia.
METHOD: Fifty healthy women having caesarean section under spinal or combined spinal-epidural anaesthesia were included. At the end of surgery, the great toe and thumb (non-i.v. fluid side) were lanced as recommended for a HemoCue reading. A venous blood sample (non-i.v. fluid side) was also taken and sent for formal laboratory measurement and tested with the HemoCue. Bland-Altman analysis was applied to the haemoglobin values.
RESULTS: Bias (mean difference) and precision +/- 2 SD were respectively 0.2 +/-1.6 for laboratory vs. toe, 0.1 +/-1.8 for laboratory vs. thumb, and 0.2 +/-1.6 laboratory vs. venous.
CONCLUSION: Our results suggest that in terms of accuracy, the two sites are equally suitable for use during caesarean section under neuraxial anaesthesia. Copyright 2009 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19945277     DOI: 10.1016/j.ijoa.2009.05.010

Source DB:  PubMed          Journal:  Int J Obstet Anesth        ISSN: 0959-289X            Impact factor:   2.603


  3 in total

1.  Comparison of the accuracy of hemoglobin point of care testing using HemoCue and GEM Premier 3000 with automated hematology analyzer in emergency room.

Authors:  Jan Zatloukal; Jiri Pouska; Jakub Kletecka; Richard Pradl; Jan Benes
Journal:  J Clin Monit Comput       Date:  2015-10-27       Impact factor: 2.502

2.  Evaluation of electrical and optical plethysmography sensors for noninvasive monitoring of hemoglobin concentration.

Authors:  Justin P Phillips; Michelle Hickey; Panayiotis A Kyriacou
Journal:  Sensors (Basel)       Date:  2012-02-09       Impact factor: 3.576

3.  Multicenter comparison of three intraoperative hemoglobin trend monitoring methods.

Authors:  Richard L Applegate Ii; Patricia M Applegate; Maxime Cannesson; Prith Peiris; Beth L Ladlie; Klaus Torp
Journal:  J Clin Monit Comput       Date:  2019-12-03       Impact factor: 2.502

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.