Yoshinori Funama1, Katsuyuki Taguchi, Kazuo Awai, Daisuke Sakabe, Masamichi Shimamura, Yasuyuki Yamashita.
Abstract
We describe the effect of the off-center position, the bowtie filter type, and the direction of scan projection radiograph on the radiation dose and image noise of 64-detector computed tomographic scans using the automatic tube current modulation technique. On anteroposterior scan projection radiographs, the tube current ratio increased, with an increase in the off-center position of the phantom irrespective of the filter size. On lateral scan projection radiographs, it remained constant with each bowtie filter type regardless of the off-center phantom position.Entities:
Mesh:
Year: 2009 PMID: 19940670 DOI: 10.1097/RCT.0b013e31819d6f6f