Image noise and radiation dose using an automatic tube current modulation technique at 64-detector computed tomography: effect of off-center patient position, bowtie filter type, and scan projection radiograph.

Yoshinori Funama1, Katsuyuki Taguchi, Kazuo Awai, Daisuke Sakabe, Masamichi Shimamura, Yasuyuki Yamashita.   

Abstract

We describe the effect of the off-center position, the bowtie filter type, and the direction of scan projection radiograph on the radiation dose and image noise of 64-detector computed tomographic scans using the automatic tube current modulation technique. On anteroposterior scan projection radiographs, the tube current ratio increased, with an increase in the off-center position of the phantom irrespective of the filter size. On lateral scan projection radiographs, it remained constant with each bowtie filter type regardless of the off-center phantom position.

Entities:  

Mesh:

Year:  2009        PMID: 19940670     DOI: 10.1097/RCT.0b013e31819d6f6f

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  2 in total

1.  Radiation dose and image quality of CT fluoroscopy with partial exposure mode.

Authors:  Keisuke Takiguchi; Atsushi Urikura; Tsukasa Yoshida; Yoshihiro Nakaya; Masahiro Endo; Takeshi Aramaki
Journal:  Diagn Interv Radiol       Date:  2020-07       Impact factor: 2.630

2.  Impact of bowtie filter and object position on the two-dimensional noise power spectrum of a clinical MDCT system.

Authors:  Daniel Gomez-Cardona; Juan Pablo Cruz-Bastida; Ke Li; Adam Budde; Jiang Hsieh; Guang-Hong Chen
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.