Literature DB >> 19940199

Pneumatic transport is critical for leukaemic patients with major leukocytosis: what precautions to measure lactate dehydrogenase, potassium and aspartate aminotransferase?

Raphael Guiheneuf1, Isabelle Vuillaume, Jacques Mangalaboyi, David Launay, Céline Berthon, J-Claude Maury, Patrice Maboudou, Jean Rousseaux.   

Abstract

False elevations of plasma lactate dehydrogenase (LDH), potassium and aspartate aminotransferase (AST) have been described, in relation to haemolysis, occurring most often by mechanical release during phlebotomy or specimen processing. We present the cases of two leukaemic patients with severe hyperleukocytosis for whom LDH, potassium and AST were dramatically but falsely elevated. This false elevation was not caused by haemolysis but could be related to white cells lysis during transport through a pneumatic transportation system, enhanced by a specific fragility of leukaemic cells. Interestingly, this interference almost completely disappeared when serum rather than plasma was used, or when leukocytosis came back to normal. This work is meant to alert clinicians to the risks of errors in LDH, potassium and AST in leukaemic patients and suggest what precautions to take.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19940199     DOI: 10.1258/acb.2009.009044

Source DB:  PubMed          Journal:  Ann Clin Biochem        ISSN: 0004-5632            Impact factor:   2.057


  3 in total

1.  Pseudohyperkalemia in patients with chronic lymphocytic leukemia.

Authors:  Stephen I Rifkin
Journal:  Int J Nephrol       Date:  2011-03-23

2.  Errors in potassium measurement: a laboratory perspective for the clinician.

Authors:  Jaya R Asirvatham; Viju Moses; Loring Bjornson
Journal:  N Am J Med Sci       Date:  2013-04

3.  Assessing Safety of Pneumatic Tube System (PTS) for Patients with Very Low Hematologic Parameters.

Authors:  Mustafa Koroglu; Mehmet Ali Erkurt; Irfan Kuku; Emin Kaya; Ilhami Berber; Ilknur Nizam; Yavuz Yagar; Seyit Ali Kayis
Journal:  Med Sci Monit       Date:  2016-04-20
  3 in total