[Evaluation of flow rate regulators for intravenous infusion].

T Caruba1, L Havard, F Gillaizeau, E Guérot, P Prognon, J Pineau.   

Abstract

OBJECTIVE: The aim of our study is to evaluate seven flow rate regulators (FRR) to assess the reliability of these devices compared to standard perfuser with roller clamp. STUDY
DESIGN: Each FRR was tested with 5% dextrose and 0.9% sodium chloride combined with three different theoretical flow rates (30, 80 and 250 ml/h). Accuracy was compared with the theoretical value. Repeatability of flow rate was assessed thanks to variance break-up.
RESULTS: Each FFR exhibits at least one combination "flow rate-solution" significantly different of the theoretical flow rate. Exadrop was the least successful of the FFR according to the accuracy. This FFR had for each combination a flow rate different of the theoretical (mean error: -24.0 ml/h). Tutodrop was the most successful of the FFR according to the accuracy with five combinations comparable to the theoretical value (mean error: -1.2 ml/h). The standard perfuser with roller clamp, used without FRR, reported two combinations comparable to the theoretical value and showed lowest rates for repeatability.
CONCLUSION: Our study exhibits the poor performances of the FRR studied: according to expected flow regulation, the reported results demonstrate the lack of accuracy. Their only one value added compare to the roller clamp is to improve the repeatability of the flow rate.

Entities:  

Mesh:

Year:  2009        PMID: 19939620     DOI: 10.1016/j.annfar.2009.09.009

Source DB:  PubMed          Journal:  Ann Fr Anesth Reanim        ISSN: 0750-7658


  1 in total

1.  The Effects of Intravenous Fluid Viscosity on the Accuracy of Intravenous Infusion Flow Regulators.

Authors:  Eunji Ko; Yeon Jae Song; Kwanyoung Choe; Yongdoo Park; Sung Yang; Choon Hak Lim
Journal:  J Korean Med Sci       Date:  2022-03-07       Impact factor: 2.153

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.