Validation of A&D UA-778 blood pressure monitor in children.

Marina V Narogan1, Marina I Narogan, Elena V Syutkina.   

Abstract

OBJECTIVE: To validate the oscillometric blood pressure monitor UA-778 in children aged 4-15 years according to the British Hypertension Society protocol.
METHOD: In 85 children (38 boys, 47 girls), nine sequential readings were taken simultaneously by two trained observers alternating between the mercury sphygmomanometer and the test device. The last seven readings were analyzed according to the British Hypertension Society protocol. The American Association for the Advancement of Medical Instrumentation criterion (mean difference +/- SD < 5 +/- 8 mmHg) was also applied.
RESULTS: The difference between UA-778 and mercury sphygmomanometer readings was 4.9 +/- 6.3 mmHg for systolic pressure and 2.4 +/- 5.9 mmHg for diastolic pressure. The proportions of device-observer blood pressure differences within 5, 10, and 15 mmHg were 50, 83, and 95% for systolic blood pressure and 70, 90, and 96% for diastolic pressure (B/A grade according to the British Hypertension Society protocol).
CONCLUSION: The oscillometric blood pressure monitor UA-778 seems to be an accurate device for blood pressure measurement in children and adolescents.

Entities:  

Mesh:

Year:  2009        PMID: 19938323     DOI: 10.1097/mbp.0b013e328330eeb2

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  3 in total

1.  Validation study of the Dinamap ProCare 200 upper arm blood pressure monitor in children and adolescents.

Authors:  Chong Guk Lee; Hyang Mi Park; Hye Jung Shin; Jin Soo Moon; Yeong Mi Hong; Nam Soo Kim; Il Soo Ha; Myeong Jin Chang; Kyeong Won Oh
Journal:  Korean J Pediatr       Date:  2011-11-30

2.  Comparison of mercury sphygmomanometry blood pressure readings with oscillometric and central blood pressure in predicting target organ damage in youth.

Authors:  Elaine M Urbina; Philip R Khoury; Connie E McCoy; Stephen R Daniels; Lawrence M Dolan; Thomas R Kimball
Journal:  Blood Press Monit       Date:  2015-06       Impact factor: 1.444

3.  Comparison of aneroid and oscillometric blood pressure measurements in children.

Authors:  Sigridur B Eliasdottir; Sandra D Steinthorsdottir; Olafur S Indridason; Runolfur Palsson; Vidar O Edvardsson
Journal:  J Clin Hypertens (Greenwich)       Date:  2013-09-20       Impact factor: 3.738

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.