Evaluation of scatter contribution and distance error by iterative methods for strength determination of HDR 192Ir brachytherapy source.

Sudhir Kumar1, Panchapakesan Srinivasan, Sunil D Sharma, Kamatam V Subbaiah, Yelia S Mayya.   

Abstract

High-dose rate (HDR) 192Ir brachytherapy sources are commonly used for management of malignancies by brachytherapy applications. Measurement of source strength at the hospital is an important dosimetry requirement. The use of 0.6-cm(3) cylindrical ionization chamber is one of the methods of measuring the source strength at the hospitals because this chamber is readily available for beam calibration and dosimetry. While using the cylindrical chamber for this purpose, it is also required to determine the positioning error of the ionization chamber, with respect to the source, commonly called a distance error (c). The contribution of scatter radiation (M(s)) from floor, walls, ceiling, and other materials available in the treatment room also need to be determined accurately so that appropriate correction can be applied while calculating the source strength from the meter reading. Iterative methods of Newton-Raphson and least-squares were used in this work to determine scatter contribution in the experimentally observed meter reading (pC/s) of a cylindrical ionization chamber. Monte Carlo simulation was also used to cross verify the results of the least-squares method. The experimentally observed, least-squares calculated and Monte Carlo estimated values of meter readings from HDR 192Ir brachytherapy source were in good agreement. Considering procedural simplicity, the method of least-squares is recommended for use at the hospitals to estimate values of f (constant of proportionality), c, and M(s) required to determine the strength of HDR 192Ir brachytherapy sources. 2010 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19931038     DOI: 10.1016/j.meddos.2009.06.008

Source DB:  PubMed          Journal:  Med Dosim        ISSN: 1873-4022            Impact factor:   1.482


  1 in total

1.  Estimation of distance error by fuzzy set theory required for strength determination of HDR (192)Ir brachytherapy sources.

Authors:  Sudhir Kumar; D Datta; S D Sharma; G Chourasiya; D A R Babu; D N Sharma
Journal:  J Med Phys       Date:  2014-04