Development of a high precision dosimetry system for the measurement of surface dose rate distribution for eye applicators.

Marion Eichmann1, Dirk Flühs, Bernhard Spaan.   

Abstract

PURPOSE: The therapeutic outcome of the therapy with ophthalmic applicators is highly dependent on the application of a sufficient dose to the tumor, whereas the dose applied to the surrounding tissue needs to be minimized. The goal for the newly developed apparatus described in this work is the determination of the individual applicator surface dose rate distribution with a high spatial resolution and a high precision in dose rate with respect to time and budget constraints especially important for clinical procedures. Inhomogeneities of the dose rate distribution can be detected and taken into consideration for the treatment planning.
METHODS: In order to achieve this, a dose rate profile as well as a surface profile of the applicator are measured and correlated with each other. An instrumental setup has been developed consisting of a plastic scintillator detector system and a newly designed apparatus for guiding the detector across the applicator surface at a constant small distance. It performs an angular movement of detector and applicator with high precision.
RESULTS: The measurements of surface dose rate distributions discussed in this work demonstrate the successful operation of the measuring setup. Measuring the surface dose rate distribution with a small distance between applicator and detector and with a high density of measuring points results in a complete and gapless coverage of the applicator surface, being capable of distinguishing small sized spots with high activities. The dosimetrical accuracy of the measurements and its analysis is sufficient (uncertainty in the dose rate in terms of absorbed dose to water is <7%), especially when taking the surgical techniques in positioning of the applicator on the eyeball into account.
CONCLUSIONS: The method developed so far allows a fully automated quality assurance of eye applicators even under clinical conditions. These measurements provide the basis for future calculation of a full 3D dose rate distribution, which then can be used as input for a refined clinical treatment planning system. The improved dose rate measurements will facilitate a clinical study, which could correlate the therapeutic outcome of a brachytherapy treatment with an applicator and its individual dose rate distribution.

Entities:  

Mesh:

Year:  2009        PMID: 19928095     DOI: 10.1118/1.3218762

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  Monte Carlo study of the energy and angular dependence of the response of plastic scintillation detectors in photon beams.

Authors:  Lilie L W Wang; David Klein; A Sam Beddar
Journal:  Med Phys       Date:  2010-10       Impact factor: 4.071

2.  Dose Distributions and Treatment Margins in Ocular Brachytherapy with 106Ru Eye Plaques.

Authors:  Erik Stöckel; Marion Eichmann; Dirk Flühs; Holger Sommer; Eva Biewald; Norbert Bornfeld; Bernhard Spaan; Wolfgang Sauerwein
Journal:  Ocul Oncol Pathol       Date:  2017-09-16

3.  Monte Carlo Estimation of Absorbed Dose Distributions Obtained from Heterogeneous 106Ru Eye Plaques.

Authors:  Francisco J Zaragoza; Marion Eichmann; Dirk Flühs; Wolfgang Sauerwein; Lorenzo Brualla
Journal:  Ocul Oncol Pathol       Date:  2017-02-23

4.  Inhomogeneous Surface Dose Distributions of 106Ru Eye Plaques.

Authors:  Marion Eichmann
Journal:  Ocul Oncol Pathol       Date:  2017-06-23

5.  Polyethylene Naphthalate Scintillator: A Novel Detector for the Dosimetry of Radioactive Ophthalmic Applicators.

Authors:  Dirk Flühs; Andrea Flühs; Melanie Ebenau; Marion Eichmann
Journal:  Ocul Oncol Pathol       Date:  2015-06-06

6.  Monte Carlo Computation of Dose-Volume Histograms in Structures at Risk of an Eye Irradiated with Heterogeneous Ruthenium-106 Plaques.

Authors:  Francisco J Zaragoza; Marion Eichmann; Dirk Flühs; Beate Timmermann; Lorenzo Brualla
Journal:  Ocul Oncol Pathol       Date:  2020-07-20

7.  Monte Carlo Simulation of the Treatment of Uveal Melanoma Using Measured Heterogeneous 106Ru Plaques.

Authors:  Francisco J Zaragoza; Marion Eichmann; Dirk Flühs; Andrea Wittig; Wolfgang Sauerwein; Lorenzo Brualla
Journal:  Ocul Oncol Pathol       Date:  2018-10-15
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.