PURPOSE: To evaluate the basic characteristics and reproducibility of anterior chamber angle (ACA) measurements determined by anterior-segment optical coherence tomography (AS-OCT) in open-angle and primary angle closure suspect (PACS) patients.
METHODS: Thirty-nine open-angle and 18 PACS patients were imaged for ACA by AS-OCT. Subjects underwent imaging of the nasal, temporal and inferior ACA under conditions of constant light, and darkness. For analysis, we used three ACA parameters handled by the Visante OCT software: angle opening distance at 500 μm (AOD(500)), trabecular-iris space area at 500 μm (TISA(500)) and angle recess area at 500 μm (ARA(500)). For determination of inter-session reproducibility, a single well-trained operator (D.Y.K.) scanned all patients at two different visits. For determination of inter-operator variability, a second operator (S.B.P.) acquired another set of images independently. Three sets of images were acquired at least 24 hour apart.
RESULTS: All parameters were significantly different when measured both in light and darkness, and in the nasal and temporal quadrants. There were no significant differences between the left and right eyes in the three ACA parameters in all quadrants. The temporal angle was wider than the nasal and inferior angles. All parameters of the nasal, temporal angles had excellent inter-session and inter-operator reproducibility [intra-class correlation coefficient (ICC) 0.796-0.981], but these values were slightly lower for inferior angle measurements (ICC 0.662-0.892) in both open-angle and PACS groups.
CONCLUSION: AS-OCT provides quantitative and reproducible assessment of ACA. Reproducibility was lower in the inferior angle compared with the nasal and temporal angles, perhaps because of variable placement of the scleral spur.
© 2009 The Authors. Journal compilation © 2009 Acta Ophthalmol.

Entities:  

Mesh:

Year:  2009        PMID: 19925516     DOI: 10.1111/j.1755-3768.2009.01714.x

Source DB:  PubMed          Journal:  Acta Ophthalmol        ISSN: 1755-375X            Impact factor:   3.761


  20 in total

1.  Comparison of the anterior ocular segment measurements using swept-source optical coherent tomography and a scanning peripheral anterior chamber depth analyzer.

Authors:  Toshie Furuya; Fumihiko Mabuchi; Tatsuya Chiba; Satoshi Kogure; Shigeo Tsukahara; Kenji Kashiwagi
Journal:  Jpn J Ophthalmol       Date:  2011-08-10       Impact factor: 2.447

2.  Assessment of the anterior chamber parameters after laser iridotomy in primary angle close suspect using Pentacam and gonioscopy.

Authors:  Alireza Esmaeili; Behzad Barazandeh; Sina Ahmadi; Alireza Haghi; Seyed Mahdi Ahmadi Hosseini; Fereshteh Abolbashari
Journal:  Int J Ophthalmol       Date:  2013-10-18       Impact factor: 1.779

3.  Residual anterior chamber angle closure in narrow-angle eyes following laser peripheral iridotomy: anterior segment optical coherence tomography quantitative study.

Authors:  Kyoung Sub Lee; Kyung Rim Sung; Sung Yong Kang; Jung Woo Cho; Dong Yoon Kim; Michael S Kook
Journal:  Jpn J Ophthalmol       Date:  2011-05-11       Impact factor: 2.447

4.  Optimal number of angle images for calculating anterior angle volume and iris volume measurements.

Authors:  Lauren S Blieden; Alice Z Chuang; Laura A Baker; Nicholas P Bell; Timothy S Fuller; Kimberly A Mankiewicz; Robert M Feldman
Journal:  Invest Ophthalmol Vis Sci       Date:  2015-05       Impact factor: 4.799

Review 5.  Test-retest variability in structural parameters measured with glaucoma imaging devices.

Authors:  Makoto Araie
Journal:  Jpn J Ophthalmol       Date:  2012-11-09       Impact factor: 2.447

6.  Pentacam could be a useful tool for evaluating and qualifying the anterior chamber morphology.

Authors:  Xiaoyu Li; Zheng Wang; Qun Cao; Lei Hu; Fang Tian; Hong Dai
Journal:  Int J Clin Exp Med       Date:  2014-07-15

7.  Evaluation of anterior chamber parameters with spectral-domain optical coherence tomography.

Authors:  Isabel Pinilla Lozano; Carmen López de la Fuente; Francisco Segura; Elvira Orduna Hospital; Ana Sánchez-Cano
Journal:  Jpn J Ophthalmol       Date:  2017-11-29       Impact factor: 2.447

8.  Identification of anterior chamber angle parameters with a portable SD-OCT device compared to a non-portable SD-OCT.

Authors:  Handan Akil; Kenneth Marion; Anna Dastiridou; Dennis Jenkins; Brendan Kramer; Brian A Francis; Vikas Chopra
Journal:  Int Ophthalmol       Date:  2016-03-24       Impact factor: 2.031

9.  Distribution of anterior chamber angle measurements in South African young adults: an optical coherence tomography study.

Authors:  Nishanee Rampersad; Rekha Hansraj
Journal:  Int Ophthalmol       Date:  2022-01-05       Impact factor: 2.029

10.  Reproducibility of scleral spur identification and angle measurements using fourier domain anterior segment optical coherence tomography.

Authors:  Ricardo J Cumba; Sunita Radhakrishnan; Nicholas P Bell; Kundandeep S Nagi; Alice Z Chuang; Shan C Lin; Kimberly A Mankiewicz; Robert M Feldman
Journal:  J Ophthalmol       Date:  2012-11-01       Impact factor: 1.909

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.