Abstract
Pure-tone thresholds were used as the reference and compared with extrapolated distortion product otoacoustic emission input/output-functions and auditory steady state responses (ASSR) in hearing-impaired adults, using the Cochlea-Scan and Audera devices. Fifty-three subjects presenting sensorineural deficits were included in the study. The DPOAE data were recorded using the detailed Cochlea-Scan threshold modality, and ASSR responses were assessed at 1.0, 2.0, and 4.0 kHz. The comparison between DPOAE and ASSR threshold values indicated significant mean differences across all tested frequencies. Significant relationships were observed between the behavioral and the DPOAE measurements in the lower frequencies (1.5 and 2.0 kHz). The Cochlea-Scan algorithm seems to overestimate hearing threshold. Logistic regression models (probability of DPOAE response p = 0.9), suggested that the identifiable hearing levels are less than 34 dB HL (at 2.0 and 4.0 kHz) and less or equal to 38 and 40 dB HL at 1.5 and 6.0 kHz respectively. The Cochlea-Scan DPOAE protocols can be used in cases presenting mild hearing deficits (i.e.<40 dB HL).Entities:
Mesh:
Year: 2009 PMID: 19925336 DOI: 10.1080/14992020902998391
Source DB: PubMed Journal: Int J Audiol ISSN: 1499-2027 Impact factor: 2.117