UNLABELLED: Aerosol delivery of Iloprost is a promising therapeutic approach. The aim of this study was to determine the output of an ultrasonic nebulizer in different ventilation set-ups at the tip of different endotracheal tubes.Entities:
Keywords: endotracheal tube; nebulization; ventilation
Year: 2009 PMID: 19920888 PMCID: PMC2761175
Source DB: PubMed Journal: Drug Des Devel Ther ISSN: 1177-8881 Impact factor: 4.162
Figure 1Set up A.
Figure 2Set up B.
Set up A. Volume-controlled ventilation (VCV)
| Run 1 [ml] | Run 2 [ml] | Run 3 [ml] | Mean [± SD] | |
|---|---|---|---|---|
| Inhalation filter | 1.58 | 1.66 | 1.60 | 1.61 ± 0.41 |
| Endotracheal tube | 0.11 | 0.12 | 0.07 | 0.10 ± 0.26 |
| Exhalation filter | 1.43 | 1.42 | 1.44 | 1.43 ± 0.01 |
| Residue in nebulizer (including all nebulizer parts) | 1.67 | 1.65 | 1.79 | 1.70 ± 0.07 |
| Nebulization time [min] | 9.0 | 12.1 | 9.3 | 10.13 ± 1.71 |
Set-up A. Pressure-controlled ventilation (PCV)
| Run 1 [ml] | Run 2 [ml] | Run 3 [ml] | Mean [± SD] | |
|---|---|---|---|---|
| Inhalation filter | 1.43 | 1.26 | 1.30 | 1.33 ± 0.88 |
| Endotracheal tube | 0.16 | 0.20 | 0.12 | 0.16 ± 0.04 |
| Exhalation filter | 1.61 | 1.42 | 1.54 | 1.52 ± 0.09 |
| Residue in nebulizer (including all nebulizer parts) | 1.73 | 2.00 | 1.95 | 1.89 ± 0.14 |
| Nebulization time [min] | 14.5 | 15.0 | 10.3 | 1.27 ± 2.58 |
Set-up B. Volume-controlled ventilation (VCV)
| Run 1 [ml] | Run 2 [ml] | Run 3 [ml] | Mean [± SD] | |
|---|---|---|---|---|
| Inhalation filter | 1.91 | 1.73 | 1.07* | 1.57 ± 0.44 |
| Endotracheal tube | 0.23 | 0.26 | 0.17 | 0.22 ± 0.05 |
| Exhalation filter | 0.21 | 0.19 | 0.72 | 0.37 ± 0.3 |
| Tube nebulizer | 0.72 | 0.8 | 0.99 | 0.84 ± 0.14 |
| Endotracheal tube | ||||
| Residue in nebulizer (including all nebulizer parts) | 1.77 | 1.98 | 1.87 | 1.87 ± 0.1 |
| Nebulization time [min] | 23.9 | 30.3 | 23.6 | 25.9 ± 3.8 |
Set-up B. Pressure-controlled ventilation (PCV)
| Run 1 [ml] | Run 2 [ml] | Run 3 [ml] | Mean [± SD] | |
|---|---|---|---|---|
| Inhalation filter | 1.22 | 1.50 | 1.18* | 1.30 ± 0.17 |
| Endotracheal tube | 0.30 | 0.24 | 0.26 | 0.27 ± 0.30 |
| Exhalation filter | 0.11 | 0.139 | 0.21 | 0.15 ± 0.05 |
| Tube nebulizer | 0.81 | 0.92 | 1.15 | 0.96 ± 0.17 |
| Endotracheal tube | ||||
| Residue in nebulizer (including all nebulizer parts) | 2.4 | 2.1 | 2.2 | 2.2 ± 0.15 |
| Nebulization time [min] | 30.1 | 22.2 | 24.4 | 25.6 ± 4.07 |
Figure 3Amount of aerosol detected in the inhalation heat and moisture exchanger.