Literature DB >> 1991767

Fatigue failure of noncemented porous-coated implants. A retrieval study.

S D Cook1, K A Thomas.   

Abstract

The causes of mechanical failure of five noncemented porous-coated components were studied. There were two cobalt-chromium alloy and three titanium alloy implants which fractured after 12 to 48 months. The implants included one acetabular component, and one femoral condylar, one patellar and two tibial components. Examination of the fractured surfaces revealed fatigue to be the mechanism of failure in all cases. The porous coating and the processes required for its fabrication had resulted in weakening and reduction of substrate thickness. Additional factors were stress concentration due to limited, localised bone ingrowth, and some features of the design of the implants.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1991767

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  6 in total

1.  A rare cause for knee pain: fracture of the femoral component after TKR. A case report.

Authors:  C Luring; L Perlick; T Schubert; M Tingart
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2006-10-06       Impact factor: 4.342

2.  Fatigue fracture of the hinge pin in a semi-constrained total knee arthroplasty: a case report.

Authors:  H R Springorum; C Luring; J Beckmann; J Schaumburger; J Grifka; M Tingart
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2010-06-19       Impact factor: 4.342

3.  Early Tibial Component Fractures in a Cementless, 3D-Printed, Titanium Implant.

Authors:  Alan D Lam; Gavan P Duffy
Journal:  Arthroplast Today       Date:  2022-10-12

4.  Fracture of the Tibial Baseplate 16 Years after Miller-Galante II Total Knee Arthroplasty.

Authors:  Kazuaki Mineta; Masahiko Okada; Soshi Matsumoto; Daisuke Hamada; Tomohiro Goto; Koichi Sairyo
Journal:  Case Rep Orthop       Date:  2017-05-16

5.  FRACTURE OF THE TIBIAL COMPONENT IN TOTAL KNEE ARTHROPLASTY: REPORT ON TWO CASES.

Authors:  Idemar Monteiro da Palma; Rodrigo Pires E Albuquerque; João Maurício Barretto
Journal:  Rev Bras Ortop       Date:  2015-12-08

6.  Tibial tray fracture in a modern prosthesis with retrieval analysis.

Authors:  John J Callaghan; David E DeMik; Nicholas A Bedard; Andrew N Odland; William M Kane; Steven M Kurtz
Journal:  Arthroplast Today       Date:  2018-02-10
  6 in total