An evaluation of the accuracy of Assess and MiniWright peak flowmeters.

S M Shapiro1, J M Hendler, R G Ogirala, T K Aldrich, M B Shapiro.   

Abstract

With the advent of small inexpensive peak flowmeters, the at-home monitoring of peak flow rates has become an invaluable aid in the treatment of asthmatic patients. In this study, we evaluated the performance of the MiniWright and Assess peak flowmeters for accuracy and reproducibility. Measurements were made at varying peak flow rates and compared with those obtained simultaneously by a calibrated pneumotachograph. When this segment of the study was completed, the peak flow devices were subjected to 200 uses and were then retested. Four MiniWright peak flowmeters that had been extensively used in our clinic were tested as well. The Assess peak flowmeter was more accurate than the MiniWright at low flow rates (less than 300 L/min), while the MiniWright meter was more accurate at high flow rates (greater than 400 L/min). We also found that the accuracy of the MiniWright meter deteriorated after 200 uses and worsened further after extensive use, while the Assess meter retained its accuracy after 200 uses.

Entities:  

Mesh:

Year:  1991        PMID: 1989795     DOI: 10.1378/chest.99.2.358

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  3 in total

1.  The accuracy of portable peak flow meters.

Authors:  M R Miller; S A Dickinson; D J Hitchings
Journal:  Thorax       Date:  1992-11       Impact factor: 9.139

2.  Peak flow meters.

Authors:  H Kawane; R Soejima; K Kojima
Journal:  Br J Gen Pract       Date:  1992-04       Impact factor: 5.386

3.  Comparison of five portable peak flow meters.

Authors:  Glaucia Nency Takara; Gualberto Ruas; Bruna Varanda Pessoa; Luciana Kawakami Jamami; Valéria Amorim Pires Di Lorenzo; Mauricio Jamami
Journal:  Clinics (Sao Paulo)       Date:  2010-05       Impact factor: 2.365

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.