| Literature DB >> 19896238 |
Samy Rengasamy1, Edward Fisher, Ronald E Shaffer.
Abstract
BACKGROUND: Respiratory protective devices exposed to pathogenic microorganisms present a potential source of transmission of infection during handling. In this study, the efficacy of 4 antimicrobial respirators to decontaminate MS2, a surrogate for pathogenic viruses, was evaluated and compared with control N95 filtering face piece respirators, which did not contain any known antimicrobial components.Entities:
Mesh:
Substances:
Year: 2009 PMID: 19896238 PMCID: PMC7124234 DOI: 10.1016/j.ajic.2009.08.006
Source DB: PubMed Journal: Am J Infect Control ISSN: 0196-6553 Impact factor: 2.918
Test antimicrobial respirators
| Respirator manufacturer | Antimicrobial component | Incorporation | Availability |
|---|---|---|---|
| A | Silver-copper based technology | Embedded into filter fiber throughout and placed on the outer layer | Prototype |
| B | EnvizO3-Shield technology | Incorporated on outer layer of respirator | Prototype |
| C | Iodinated resin | Incorporated on filtering layer and positioned within respirator | Commercially available |
| D | TiO2 | Coated on filtering layers and placed beneath the outer layer | Prototype |
MS2 loading on to N95 filtering face piece respirators and antimicrobial respirators
| Antimicrobial respirator | |||||
|---|---|---|---|---|---|
| MS2 loading | N95 FFR | A | B | C | D |
| Set 1: loading (log10) | 5.96 ± 0.18 (3%) | 5.93 ± 0.2 (3%) | 5.67 ± 0.07 (1%) | 6.19 ± 0.27 (4%) | 6.38 ± 0.07 (1%) |
| Set 2: loading (log10) | 5.71 ± 0.31 (5%) | 6.50 ± 0.15 (2%) | 5.13 ± 0.33 (6%) | 5.81 ± 0.36 (6%) | ND |
NOTE. Numbers in parentheses are relative standard deviations of the loading levels for N95 FFRs and antimicrobial respirators.
Fig 1MS2 loaded respirator coupons were incubated at 22°C and 30% relative humidity (RH), and viable MS2 were enumerated at 8 and 20 hours.
Fig 2MS2 loaded respirator coupons were incubated at 37°C and 80% relative humidity (RH), and viable MS2 were enumerated at 2 and 4 hours. ∗Significantly (P < .05) different from the control N95 respirator.