Literature DB >> 19893856

Development of a system for measurement and analysis of tremor using a three-axis accelerometer.

N Mamorita1, T Iizuka, A Takeuchi, M Shirataka, N Ikeda.   

Abstract

OBJECTIVES: The aim of the study was to develop a low-cost and compact system for analysis of tremor using a three-axis accelerometer (the Wii Remote (Nintendo)). To analyze tremor, we hypothesized that the influence of gravitational acceleration should be separated from that of movement. This hypothesis was tested experimentally and we also attempted to record and analyze tremor using our system in a clinical ward.
METHODS: A system for tremor measurement and analysis was developed using the three-axis accelerometer built into the Wii Remote. The frequency and amplitude of mechanical oscillation were calculated using methods for frequency analysis of the axis of largest variance and an estimation of tremor amplitude.
RESULTS: The system consists of a program for measurement and analysis of Wii Remote acceleration (Tremor Analyzer), a Wii Remote, a Bluetooth USB adapter and a Web camera. The Tremor Analyzer has a GUI (graphical user interface) that is divided into five seg- ments. The sampling period of the analyzer is 30 msec. To confirm the hypothesis, mechanical oscillations were fed to the Wii Remote. The peak frequency of the power spectrum and the frequency of the oscillation generator were in good agreement, except at 1 Hz (0.01 G) and 2 Hz (0.02 G). With a change in the sum of squares of the three axes from 1.0 to 1.8 (G), the estimated and generated amplitude (0.3 cm) were in close agreement.
CONCLUSIONS: This system using a Wii Remote is capable of analyzing frequency and estimated amplitude of tremor between 3 Hz and 15 Hz.

Entities:  

Mesh:

Year:  2009        PMID: 19893856     DOI: 10.3414/ME9243

Source DB:  PubMed          Journal:  Methods Inf Med        ISSN: 0026-1270            Impact factor:   2.176


  4 in total

Review 1.  Biomarker candidates of neurodegeneration in Parkinson's disease for the evaluation of disease-modifying therapeutics.

Authors:  Manfred Gerlach; Walter Maetzler; Karl Broich; Harald Hampel; Lucas Rems; Torsten Reum; Peter Riederer; Albrecht Stöffler; Johannes Streffer; Daniela Berg
Journal:  J Neural Transm (Vienna)       Date:  2011-07-14       Impact factor: 3.575

2.  Continuous Monitoring of Essential Tremor Using a Portable System Based on Smartwatch.

Authors:  Xiaochen Zheng; Alba Vieira Campos; Joaquín Ordieres-Meré; Jose Balseiro; Sergio Labrador Marcos; Yolanda Aladro
Journal:  Front Neurol       Date:  2017-03-15       Impact factor: 4.003

3.  Wearable sensors during drawing tasks to measure the severity of essential tremor.

Authors:  Sheik Mohammed Ali; Sridhar Poosapadi Arjunan; James Peters; Laura Perju-Dumbrava; Catherine Ding; Michael Eller; Sanjay Raghav; Peter Kempster; Mohammod Abdul Motin; P J Radcliffe; Dinesh Kant Kumar
Journal:  Sci Rep       Date:  2022-03-28       Impact factor: 4.379

4.  Feasibility of a New Desktop Motion Analysis System with a Video Game Console for Assessing Various Three-Dimensional Wrist Motions.

Authors:  Kwang Gi Kim; Chan Soo Park; Suk Ha Jeon; Eui Yub Jung; Jiyun Ha; Sanglim Lee
Journal:  Clin Orthop Surg       Date:  2018-11-21
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.