Pradeep Narayan1, S G Vaijapurkar, P K Bhatnagar.
Abstract
A commercial glass plate of thickness 1.75 mm has been utilized for evaluation of dose distributions inside the irradiation volume of gamma chamber using optical densitometry technique. The glass plate showed linear response in the dose range 0.10 Kilo Gray (kGy) to 10 kGy of cobalt-60 gamma radiation with optical sensitivity 0.04 Optical Density (OD) /kGy. The change in the optical density at each identified spatial dose matrix on the glass plate in relation to the position in the irradiation volume has been presented as dose distributions inside the gamma chamber. The optical density changes have been graphically plotted in the form of surface diagram of color washes for different percentage dose rate levels as isodose distributions in gamma chamber. The variation in dose distribution inside the gamma chamber unit, GC 900, BRIT India make, using this technique has been observed within +/- 15%. This technique can be used for routine quality assurances and dose distribution validation of any gamma chamber during commissioning and source replacement. The application of commercial glass plate for dose mapping in gamma chambers has been found very promising due to its wider dose linearity, quick measurement, and lesser expertise requirement in application of the technique.Entities:
Keywords: Dose distribution; flat glass plate; gamma chamber; high dose measurement; optical densitometry
Year: 2008 PMID: 19893695 PMCID: PMC2772032 DOI: 10.4103/0971-6203.41277
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Calibration of glass detector (0.10 kGy-10.00 kGy) against cobalt-60 gamma radiation
| 100 | 0.02 | 0.00 | 0.00 |
| 350 | 0.03 | 0.00 | 0.00 |
| 600 | 0.04 | 0.00 | 0.00 |
| 850 | 0.05 | 0.00 | 0.00 |
| 3050 | 0.13 | 0.00 | 0.00 |
| 5900 | 0.23 | 0.01 | 4.35 |
| 8850 | 0.33 | 0.01 | 3.03 |
| 10000 | 0.36 | 0.01 | 2.78 |
Figure 1Optical density changes of the glass plate exposed with cobalt-60 gamma radiation (100 Gy-10 kGy)
Figure 2Dose distribution in the irradiation volume of gamma chamber (GC 900), BRIT India, using glass plate exposed with 5 kGy of cobalt-60 gamma radiation
Figure 3Variation of dose rate along the diameter of cylindrical irradiation volume of gamma chamber (GC 900), BRIT India
Figure 4Variation of dose rate along the central axis of cylindrical irradiation volume of gamma chamber (GC 900), BRIT India