Quantitation of humalog insulin by reversed-phase high-performance liquid chromatography.

Carey Munsick1, Robert Murray, Tom Dziubla, Anthony M Lowman, Jeffrey I Joseph, Marc C Torjman.   

Abstract

INTRODUCTION: The methodology for in vitro testing of insulin delivery systems for long-term infusion of lispro insulin requires insulin flow studies over time, with the measurement of lispro concentrations in wells or other laboratory fluid collection systems. We postulated that the efficiency of the insulin assay could be improved if the insulin collected in the wells could be measured without having to perform dilutions of insulin samples. The manufacturer's method for the identification of Humalog(R) insulin by high-performance liquid chromatography (HPLC) does not provide a detailed method for the quantitation of the clinical formulation. No reports could be found in the literature describing a method to quantitate lispro insulin's entire concentration range, nor could we find detailed information on pH effects. The purpose of this study was to investigate the use of a reversed-phase HPLC method to quantitate Humalog insulin at three levels of pH.
METHODS: Serial dilutions of Humalog insulin stock solution were prepared at pH levels of 2.6 and 7.4 (concentrations 0.20-100 IU/ml) to construct calibration curves. Samples were also prepared at lower concentrations (0.23-15.0 IU/ml) at pH levels of 4.0 and 7.4. Areas under curve data were plotted against known concentrations, and the limit of quantitation and precision of the assay were determined.
RESULTS: Calibration curves for the 0.20-100 IU/ml concentration (pH 7.4 and 2.6) revealed no significant differences (p < 0.01) between those curves. pH and concentration did not have an effect on the overall precision of the method. Linearity was not conserved past the 15-IU/ml concentration. Coefficient of variation (CV) values were generally <15% except at the lower concentrations where the largest CVs were observed. Mean known concentrations plotted on a logarithmic scale demonstrated linearity from 12.5 to 100 IU/ml.
CONCLUSIONS: Humalog insulin clinical formulation can be quantitated over the entire concentration range by a reversed-phase HPLC method using nonlinear regression analysis. The method is reproducible at lower (<15 IU/ml) and higher insulin concentrations. Linear regression analysis may be used when the concentrations of interest are in the 0- to 15-IU range. Preparation of insulin solutions at pH 2.6, 4.0, and 7.4 did not significantly affect the reproducibility of the assay.

Entities:  

Keywords:  HPLC assay; Humalog; drug delivery systems; insulin; lispro

Year:  2007        PMID: 19885125      PMCID: PMC2769640          DOI: 10.1177/193229680700100420

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  11 in total

1.  Assessment of in vivo stability of a new insulin preparation for implantable insulin pumps. A randomized multicenter prospective trial. EVADIAC Group. Evaluation Dans le diabète du Traitement par Implants Actifs.

Authors:  S Boivin; P Belicar; V Melki
Journal:  Diabetes Care       Date:  1999-12       Impact factor: 19.112

2.  Improved stability of insulin delivery from implanted pumps using a new preparation process for infused insulin.

Authors:  E Renard; C Souche; D Jacques-Apostol; D Lauton; F Gibert-Boulet; G Costalat; J Bringer; C Jaffiol
Journal:  Diabetes Care       Date:  1999-08       Impact factor: 19.112

3.  Metabolic decompensation in pump users due to lispro insulin precipitation.

Authors:  Howard A Wolpert; Raquel N Faradji; Susan Bonner-Weir; Myra A Lipes
Journal:  BMJ       Date:  2002-05-25

4.  Precipitation of insulin products used for continuous subcutaneous insulin infusion.

Authors:  Christian Poulsen; Lotte Langkjaer; Carsten Worsøe
Journal:  Diabetes Technol Ther       Date:  2005-02       Impact factor: 6.118

5.  Stability of insulin lispro in insulin infusion systems.

Authors:  W D Lougheed; B Zinman; T R Strack; L J Janis; A B Weymouth; E A Bernstein; A M Korbas; B H Frank
Journal:  Diabetes Care       Date:  1997-07       Impact factor: 19.112

6.  Cannula occlusion with use of insulin lispro and insulin infusion system.

Authors:  A W Wright; J A Little
Journal:  Diabetes Care       Date:  1998-05       Impact factor: 19.112

7.  Lispro insulin is suitable for external pumps but not for implantable pumps.

Authors:  S Demirdjian; C Bardin; S Savin; P Zirinis; F Chast; G Slama; J L Sélam
Journal:  Diabetes Care       Date:  1998-05       Impact factor: 19.112

8.  In vitro stability of insulin lispro in continuous subcutaneous insulin infusion.

Authors:  Michael R DeFelippis; Maureen A Bell; Jennifer A Heyob; Sacha M Storms
Journal:  Diabetes Technol Ther       Date:  2006-06       Impact factor: 6.118

Review 9.  Insulin aggregation in artificial delivery systems.

Authors:  W D Lougheed; H Woulfe-Flanagan; J R Clement; A M Albisser
Journal:  Diabetologia       Date:  1980-07       Impact factor: 10.122

10.  A method for assessing catheter patency in implanted pumps for long-term intraperitoneal insulin delivery.

Authors:  M Scavini; B Hammarberg; F Dosio; M Torri; G Petrella; G Galimberti; S Vai; F Fazio; G Pozza
Journal:  Artif Organs       Date:  1997-05       Impact factor: 3.094

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.