Comparison of conventional and torsional mode (OZil) phacoemulsification: randomized prospective clinical study.

Ercument Bozkurt1, Sukru Bayraktar, Serpil Yazgan, Mehmet Cakir, Osman Cekic, Havva Erdogan, Omer F Yilmaz.   

Abstract

PURPOSE: To compare the amount of intraoperative energy used during conventional and torsional phacoemulsification surgery and investigate the changes in corneal endothelial cell density.
METHODS: In this study, a total of 100 eyes in patients with age-related cataract (mean age 68.9 years) were included. Patients were randomly allocated into conventional (group 1) or torsional (OZil, group 2) phacoemulsification. Mean nuclear grade made according to Lens Opacities Classification System II was not different in the 2 groups. Total ultrasound time (USTT), cumulative dissipated energy (CDE), and the percent of total equivalent power in position 3 (%USTEPiP3) were compared between the 2 groups.
RESULTS: There were 47 eyes in group 1 and 53 eyes in group 2. Patient age and sex were matched between the 2 groups (p=0.49, p=0.08). Mean USTT was 1.6+/-1.1 minutes (m) in group 1 and 1.5 +/- 0.9 m in group 2 (p=0.55). Average CDE was 29.9+/-16.9 in group 1 and 25.2+/-19.1 in group 2 (p=0.20).Mean %USTEPiP3 was 27.0%+/-8.5% in group 1; however, it was 22.4%+/-6.6% in group 2 (p=0.003). At the first postoperative day, mean corneal edema was graded as 0.25+/-0.5 in group 1 and 0.18 +/- 0.4 in group 2 (p=0.47). At the first postoperative week, it was found to be 0.02+/-0.15 in group 1 and 0.03+/-0.27 in group 2 (p=0.71). ECD loss was 6.7%+/-3.3% in group 1 and 4.2%+/-5.7% in group 2 (p=0.56).
CONCLUSION: No statistically significant difference was found between conventional and torsional phacoemulsification with respect to USTT and CDE. Torsional mode led to a significant reduction in US power in footswitch position 3. Torsional mode also appears to be associated with less loss of ECD.

Entities:  

Mesh:

Year:  2009        PMID: 19882578     DOI: 10.1177/112067210901900614

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   1.922


  8 in total

1.  Ozil® versus conventional ultrasound phacoemulsification: a randomized comparative study.

Authors:  Afef Maalej; Asma Khallouli; Cheima Wathek; Riadh Rannen
Journal:  Int Ophthalmol       Date:  2014-08-26       Impact factor: 2.031

2.  Comparative clinical study of Whitestar Signature phacoemulsification system with standard and Ellips FX handpieces.

Authors:  Daniele Tognetto; Rossella D'Aloisio; Paolo Cecchini; Marta Di Nicola; Giuseppe Di Martino
Journal:  Int Ophthalmol       Date:  2017-07-13       Impact factor: 2.031

3.  Comparison of corneal endothelial changes following phacoemulsification with transversal and torsional phacoemulsification machines.

Authors:  Mustafa Ataş; Süleyman Demircan; Arzu Seyhan Karatepe Haşhaş; Ahmet Gülhan; Gökmen Zararsız
Journal:  Int J Ophthalmol       Date:  2014-10-18       Impact factor: 1.779

4.  An evaluation of intraoperative and postoperative outcomes of torsional mode versus longitudinal ultrasound mode phacoemulsification: a Meta-analysis.

Authors:  Pia Leon; Ingrid Umari; Alessandro Mangogna; Andrea Zanei; Daniele Tognetto
Journal:  Int J Ophthalmol       Date:  2016-06-18       Impact factor: 1.779

5.  Torsional ultrasound mode versus combined torsional and conventional ultrasound mode phacoemulsification for eyes with hard cataract.

Authors:  Mohamed A Fakhry; Malak I El Shazly
Journal:  Clin Ophthalmol       Date:  2011-07-15

6.  Comparison of Femtosecond Laser-Assisted Cataract Surgery and Conventional Phacoemulsification on Endothelial Cell Density When Using Torsional Modality.

Authors:  Yoshitaka Oka; Noriyuki Sasaki; Valentine P Injev
Journal:  Clin Ophthalmol       Date:  2021-10-20

7.  Morphological variations influencing the outcomes in posterior polar cataract.

Authors:  P R Aswin; Kandukuri Harika; Madhu Shekhar; R Sankarananthan; Amish Shah; P Lakshmanan; Haripriya Aravind
Journal:  Indian J Ophthalmol       Date:  2022-07       Impact factor: 2.969

8.  Comparative studies between longitudinal and torsional modes in phacoemulsification, using active fluidics technology along with the intrepid balanced tip.

Authors:  Sushobhan Dasgupta; Rohan Mehra
Journal:  Indian J Ophthalmol       Date:  2018-10       Impact factor: 1.848