Literature DB >> 19881085

Technical note: A simple and effective CO 2 delivery system for angiography using a blood bag.

Mathew P Cherian1, Pankaj Mehta, Prashant Gupta, Tejas M Kalyanpur, S R Jayesh, R Rupa.   

Abstract

Several angiographic techniques have been developed to image the arterial system, the commonest using iodinated contrast media. Useful as they may be, they are not without disadvantages. One other modality is angiography using CO 2 . Although CO 2 can be used as an alternative contrast medium, delivery systems are expensive to procure. We describe an indigenous and effective delivery system developed at our institute.

Entities:  

Year:  2009        PMID: 19881085      PMCID: PMC2766871          DOI: 10.4103/0971-3026.54881

Source DB:  PubMed          Journal:  Indian J Radiol Imaging        ISSN: 0970-2016


Introduction

Vascular diseases of the arterial system have long been a major cause of morbidity and mortality. Although several noninvasive techniques have evolved in recent times to image the arterial system, the gold standard remains the catheter digital subtraction angiography (DSA). Though the iodinated contrast media used today are relatively safe, they should, however, be used with caution in patients with compromised renal function and in diabetics. Patients with inadequate cardiac function are also poor candidates for angiography with iodinated contrast. Carbon dioxide (CO2) has a role in this subset of patients as an effective and safe contrast agent.[1] Though gadolinium may be used as a substitute contrast agent[1] in patients with chronic renal failure, there is a chance of developing a potentially fatal complication called nephrogenic systemic fibrosis.[23] CO2 has received immense attention in recent times for its excellent safety profile. There is a paucity of studies, related to procedures using CO2 in the Indian medical literature; a PubMed search in fact yielded no study from India regarding the use of CO2 in vascular procedures; this is probably due to the fact that dedicated delivery systems for CO2 are expensive to procure and maintain. This problem has been solved by developing an indigenous injection system at our institute.

Materials and Methods

The system developed at our institute has been used for CO2 angiograms for many years now. It consists of a standard blood bag (readily available) connected to a CO2 cylinder [Figure 1]; this is done in order to avoid direct transmission of pressure from the cylinder to the vasculature under examination. The blood bag is then connected to a three-way stopcock [Figure 2], to which are connected a 50-cc plastic syringe and the angiographic catheter to be used [Figure 3].
Figure 1

The blood bag inflated with CO2 is depicted

Figure 2

The syringe, three-way, and catheter assembly are shown

Figure 3

This schematic shows our delivery system

The blood bag inflated with CO2 is depicted The syringe, three-way, and catheter assembly are shown This schematic shows our delivery system The blood bag is first filled with CO2 and the knob on the cylinder is closed. The bag now acts as a CO2 reservoir. CO2 is then allowed to freely flow for some time through the three ways in a basin containing sterile saline. This ensures that the tubings are free of air. The blood bag is intermittently filled to ensure that there is adequate CO2 in this low pressure reservoir. With the CO2 flowing through the ports of the three ways, a luer lock syringe is connected ensuring that the syring is kept upside down. A 50 cc syringe is used if the injection has to be made in the renal or infra renal aorta. We would use a 20 cc syringe in the iliac and femoral arteries. Seventy five percent of the syringe is filled with CO2 by gentle aspiration. The catheter is flushed with about 3 cc of CO2 to prevent explosive delivery of the agent. According to Caridi et al,[1] this explosive delivery is a major cause of patient discomfort. Once the catheter has been flushed, fifty percent of the CO2 in the syringe is injected to adequately visualize the vessel. The CO2 displaces blood in the vessel and provides ‘negative contrast’ [Figure 4]. The image quality can be enhanced by using ‘stacking software’. Stacking software is used because the CO2 column undergoes fragmentation as it travels distally; the software post-processes the images by stacking them one over the other, thus leading to an image in which the entire CO2 column is visualized (minimum intensity stacking software). This injection system was employed in a prospective study between September 2005 and August 2007, wherein 44 patients underwent lower limb arterial angiograms with CO2 and iodinated contrast (IC).
Figure 4

Angiogram with CO2 using stacking software shows the superficial femoral artery (arrow)

Angiogram with CO2 using stacking software shows the superficial femoral artery (arrow)

Results

Of the 44 cases evaluated with this system, we obtained diagnostic quality angiograms in 42 (95.5%) cases. The system was easy to use and required minimal manpower and expertise to maintain and setup. There was no system malfunction at any time and no complications were encountered.

Discussion

The commercially available delivery systems are not readily available in India and are difficult and expensive to procure. Hawkins et al,[4] developed a plastic bag system in 1995, which they later modified[5] to include O-rings to reduce the possibility of a CO2 leak; this system is commercially available as the Angioflush III. Our system has been indigenously developed and uses commonly available materials that are cheap to procure and maintain. The use of the blood bag as the reservoir avoids transmission of the high pressure of the gas in the cylinder to the vasculature; it also allows us to fill the syringe with a limited quantity of CO2. We circumvented the problem of CO2 leaks by using a three-way stopcock to which the catheter, syringe and the blood bag are attached. With proper operation of the stopcock, room air contamination can be completely avoided. This is an effective and cheap alternative to the commercially available CO2 delivery systems.
  5 in total

1.  Modified plastic bag system with O-ring fitting connection for carbon dioxide angiography.

Authors:  I F Hawkins; J G Caridi; S D Klioze; C R Mladinich
Journal:  AJR Am J Roentgenol       Date:  2001-01       Impact factor: 3.959

2.  Is there a causal relation between the administration of gadolinium based contrast media and the development of nephrogenic systemic fibrosis (NSF)?

Authors:  H S Thomsen; S K Morcos; P Dawson
Journal:  Clin Radiol       Date:  2006-11       Impact factor: 2.350

3.  Plastic bag delivery system for hand injection of carbon dioxide.

Authors:  I F Hawkins; J G Caridi; S R Kerns
Journal:  AJR Am J Roentgenol       Date:  1995-12       Impact factor: 3.959

Review 4.  Carbon dioxide digital subtraction angiography: the practical approach.

Authors:  J G Caridi; I F Hawkins; S D Klioze; R F Leveen
Journal:  Tech Vasc Interv Radiol       Date:  2001-03

5.  Gadolinium is detectable within the tissue of patients with nephrogenic systemic fibrosis.

Authors:  Whitney A High; Reed A Ayers; John Chandler; Gary Zito; Shawn E Cowper
Journal:  J Am Acad Dermatol       Date:  2006-11-09       Impact factor: 11.527

  5 in total
  1 in total

1.  Carbon dioxide Angiography-Guided Renal-Related Interventions in Patients with Takayasu Arteritis and Renal Insufficiency.

Authors:  Sujith Chacko; George Joseph; Viji Thomson; Paul George; Oommen George; Debashish Danda
Journal:  Cardiovasc Intervent Radiol       Date:  2018-03-16       Impact factor: 2.740

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.