Anterior chamber volume measurements with Visante optical coherence tomography and Pentacam: repeatability and level of agreement.

Georgios Labiris1, Maria Gkika, Andreas Katsanos, Michael Fanariotis, Efstratios Alvanos, Vassilios Kozobolis.   

Abstract

BACKGROUND: The purpose of this study is to examine the repeatability of Pentacam-derived anterior chamber volume measurements, and assess the level of agreement of anterior chamber volume measurements obtained with the Visante and the Pentacam imaging devices.
METHODS: Anterior chamber volume was calculated with the Pentacam device and the Visante using the Wang formula. Results were analysed by means of the Wilcoxon signed rank test, Bland-Altman analysis, Cronbach's alpha test and intraclass correlation coefficient. P-values <0.01 were considered statistically significant.
RESULTS: One randomly selected eye from each of 20 persons was analysed. Spherical equivalent ranged between -2.50 and +1.50 D, and their age ranged from 25 to 79 years (mean +/- SD: 54.4 +/- 19.2 years). Mean anterior chamber volume measured with Pentacam was 171.1 +/- 39.6 microL (minimum: 103.0 microL, maximum: 236.0 microL), and mean anterior chamber volume measured by Visante was 171.4 +/- 42.4 microL (minimum: 96.4 microL, maximum: 245.5 microL). There was no statistically significant difference between Pentacam and Visante measurements of the anterior chamber volume (P = 0.691, Wilcoxon signed rank test). Cronbach's alpha and intraclass correlation coefficient for Pentacam were 0.998 and 0.991, respectively.
CONCLUSIONS: Anterior chamber volume measurements obtained with the Visante using the Wang formula were in good agreement with the corresponding ones obtained with Pentacam. This recently described method is easy to perform and might contribute to both clinical and research settings.

Entities:  

Mesh:

Year:  2009        PMID: 19878221     DOI: 10.1111/j.1442-9071.2009.02132.x

Source DB:  PubMed          Journal:  Clin Exp Ophthalmol        ISSN: 1442-6404            Impact factor:   4.207


  5 in total

1.  Air entry into the anterior chamber post intravitreal injection of Eylea.

Authors:  Wei Sing Lim; Munir Sikandar; Heather Jackson
Journal:  BMJ Case Rep       Date:  2016-07-20

2.  Clinical intraocular islet transplantation is not a number issue.

Authors:  A Shishido; A Caicedo; R Rodriguez-Diaz; A Pileggi; P-O Berggren; M H Abdulreda
Journal:  CellR4 Repair Replace Regen Reprogram       Date:  2016-07-26

3.  Comparison of Scheimpflug imaging and spectral domain anterior segment optical coherence tomography for detection of narrow anterior chamber angles.

Authors:  D S Grewal; G S Brar; R Jain; S P S Grewal
Journal:  Eye (Lond)       Date:  2011-02-18       Impact factor: 3.775

4.  Anterior segment Scheimpflug imaging for detecting primary angle closure disease.

Authors:  Andrew Winegarner; Atsuya Miki; Miho Kumoi; Yuichiro Ishida; Taku Wakabayashi; Susumu Sakimoto; Shinichi Usui; Kenji Matsushita; Kohji Nishida
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2018-10-30       Impact factor: 3.117

5.  No change in anterior chamber dimensions after femtosecond LASIK for hyperopia.

Authors:  Xiaodong Zhou; Tao Li; Zhi Chen; Lingling Niu; Xingtao Zhou; Zimei Zhou
Journal:  Eye Contact Lens       Date:  2015-05       Impact factor: 2.018

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.