Closed-circuit xenon delivery using a standard anesthesia workstation.

Shilpa Rawat1, John Dingley.   

Abstract

BACKGROUND: Xenon (Xe) is an anesthetic with minimal side effects, now also showing promise as a neuroprotectant both in vitro and in vivo. Although scarce and expensive, Xe is insoluble and patient uptake is low, making closed circuits the optimum delivery method. Although the future of Xe anesthesia is uncertain, effective neuroprotection is highly desirable even if moderately expensive. A factor limiting Xe research in all these fields may be the perceived need to purchase special Xe anesthesia workstations that are expensive and difficult to service. We investigated the practicality of 1) true closed-circuit Xe delivery using an unmodified anesthesia workstation with gas monitoring/delivery attachments restricted to breathing hoses only, 2) a Xe delivery protocol designed to eliminate wastage, and 3) recovering Xe from exhaled gas.
METHODS: Sixteen ASA physical status I/II patients were recruited for surgery of > 2 h. Denitrogenation with 100% oxygen was started during induction and tracheal intubation under propofol/remifentanil anesthesia. This continued after operating room transfer for 30 min. All fresh gases were then temporarily stopped, metabolic oxygen consumption then being replaced with 250-mL Xe boluses until F(I)Xe = 50%. A basal oxygen fresh gas flow was thereafter restored with additional Xe given as required via the expiratory hose to maintain a F(I)Xe > or = 50%. At no time, apart from during circle flushes every 90 min, were the bellows allowed to completely fill and spill gas, ensuring the circle remained closed. On termination of anesthesia, the first 10 exhaled breaths were collected as was residual gas from the circle, allowing measurement of the Xe content of each.
RESULTS: Total Xe consumption, including initial wash-in and circle flushes, was 12.62 (5.31) L or 4.95 (0.82) L/h, mean (sd). However, consumption during maintenance periods was lower: 3 L/h at 1 h and 2 L/h thereafter. Of the total Xe used, 8.98% (5.94%) could be recovered at the end of the procedure.
CONCLUSIONS: We report that closed-circuit Xe delivery can be achieved with a modified standard anesthesia workstation with breathing hose alterations only and that the protocol was very gas efficient, especially during the normally wasteful Xe wash-in. A Xe mixture of > or = 50% was delivered for up to 341 min (5 h 41 min) and Xe consumption was 4.95 (0.82) L/h, maintenance being achieved with 2-3 L/h. With this degree of efficiency, Xe recovery/recycling at the end of anesthesia may be of little additional benefit.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19861365     DOI: 10.1213/ANE.0b013e3181be0e17

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  9 in total

Review 1.  Noble gases as cardioprotectants - translatability and mechanism.

Authors:  Kirsten F Smit; Nina C Weber; Markus W Hollmann; Benedikt Preckel
Journal:  Br J Pharmacol       Date:  2015-01-12       Impact factor: 8.739

Review 2.  Waste anesthetic gas exposure and strategies for solution.

Authors:  Hai-Bo Deng; Feng-Xian Li; Ye-Hua Cai; Shi-Yuan Xu
Journal:  J Anesth       Date:  2018-02-05       Impact factor: 2.078

3.  A physiologically based model for denitrogenation kinetics.

Authors:  Ira Katz; Jacqueline Murdock; Marc Palgen; Géraldine Farjot
Journal:  Med Gas Res       Date:  2018-01-22

4.  Pharmacokinetic analysis of the chronic administration of the inert gases Xe and Ar using a physiological based model.

Authors:  Ira Katz; Jacqueline Murdock; Marc Palgen; Jan Pype; Georges Caillibotte
Journal:  Med Gas Res       Date:  2015-05-29

5.  The Modification and Performance of a Large Animal Anesthesia Machine (Tafonius®) in Order to Deliver Xenon to a Horse.

Authors:  Bruna Santangelo; Astrid Robin; Keith Simpson; Julie Potier; Michel Guichardant; Karine Portier
Journal:  Front Vet Sci       Date:  2017-09-29

Review 6.  Potential application value of xenon in stroke treatment.

Authors:  Chong-Shun Zhao; Hao Li; Zhong Wang; Gang Chen
Journal:  Med Gas Res       Date:  2018-09-25

7.  Xenon consumption during general surgery: a retrospective observational study.

Authors:  Christian Stoppe; Achim Rimek; Rolf Rossaint; Steffen Rex; Ana Stevanovic; Gereon Schälte; Astrid Fahlenkamp; Michael Czaplik; Christian S Bruells; Christian Daviet; Mark Coburn
Journal:  Med Gas Res       Date:  2013-06-11

8.  Xenon treatment protects against cold ischemia associated delayed graft function and prolongs graft survival in rats.

Authors:  H Zhao; H R Watts; M Chong; H Huang; C Tralau-Stewart; P H Maxwell; M Maze; A J T George; D Ma
Journal:  Am J Transplant       Date:  2013-05-24       Impact factor: 8.086

9.  Closed circuit xenon delivery for 72h in neonatal piglets following hypoxic insult using an ambient pressure automated control system: Development, technical evaluation and pulmonary effects.

Authors:  John Dingley; Satomi Okano; Richard Lee-Kelland; Emma Scull-Brown; Marianne Thoresen; Ela Chakkarapani
Journal:  PLoS One       Date:  2020-01-21       Impact factor: 3.240

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.